3
Device overview
36 mm
2 threaded holes
for clamping block
M6 x 9 mm
Waterproof RJ45 connector
acc. to IEC 61076-3-106,
variant 4
LED
Protective cap
Base plate
106.2 mm
Ø 115.5 mm
14.5 mm
Threaded hole
1/4" BSW x 6.5 mm
for temporary mounting
on a stand
Membrane vent covering the pushbutton
(tightening torque: 0.6 - 0.8 Nm)
Type plate with
product name
and serial number
Degree of protection:
IP67 (for outdoor use) if the waterproof
RJ45 connector is inserted properly
Weight: < 500 g
OTMC 100
does not provide an ON/OFF switch. The device automatically powers up after supply
voltage is provided via Ethernet (PoE).
The
LED
indicates the device status:
LED continuously off
OTMC 100
is not supplied with power via the RJ45 Ethernet
connector.
LED lights red
Boot process during power-up or device reboot is in progress.
LED flashes red
Software update is in progress.
ATTENTION: Do not disconnect
OTMC 100
from its PoE source
(Ethernet network) during a software update!
LED lights orange
Intermediate state when initiating a factory reset (see also
procedures performed directly on the device
LED flashes green
OTMC 100
is in the recovery mode, waiting for new software. In the
recovery mode, the device provides only a rudimentary web interface
just allowing for the upload of a software image.
LED lights green
OTMC 100
is ready for operation.
The green LED does not provide information about the
number of satellites received or the actual time accuracy
provided by
OTMC 100
. Such information are displayed in
the
Status
The
pushbutton
can be accessed after unscrewing the water-tight membrane vent.