Error and Warning Messages
12
Error and Warning Messages
001.xxx
Error [001] No valid CT software!
Reason:
No valid software in the device Flash memory or incorrect checksum.
Solution:
Insert a Compact Flash card with valid software (
CTAnalyzer.bin
) in the
Omicron
directory and switch the
CT Analyzer
off and on again.
002.xxx
Error [002] Can't open file!
Reason:
The firmware cannot read the file
CTAnalyzer.bin
from the Compact Flash
card since the CF card or the file is missing.
Solution:
Insert a Compact Flash card with valid software (
CTAnalyzer.bin
) in the
Omicron
directory and switch the
CT Analyzer
off and on again.
003.xxx
Error [003] Download error!
Reason:
The downloaded software is corrupt.
Solution:
Insert a Compact Flash card with valid software (
CTAnalyzer.bin
) in the
Omicron
directory and switch the
CT Analyzer
off and on again.
100.xxx
Warning [100.xxx] CT resistance > 3000 Ohms.
Reason:
The resistance measured during primary or secondary winding resistance
measurement is > 3000
Ω
.
Solution:
The test cannot be continued. Check the connections to the CT. If the
connections are ok, the CT resistance is > 3000
Ω
. Such CTs cannot be tested
using the
CT Analyzer
.
101.xxx
Warning [101.xxx] Timeout during measurement. No constant winding resistance can
be determined within 10 sec.
Reason:
The
CT Analyzer
cannot determine a constant winding resistance within
10sec. during primary or secondary winding resistance measurement.
Solution:
The test cannot be continued. Check the connections to the CT.
110.xxx
Warning [110.xxx] The impedance of the CT is too high.
Reason:
The impedance of the CT or the load is too high to reach the minimum toggle
frequency of 0.2Hz / 0.8Hz.
Solution:
The test cannot be continued. Check the connections to the CT. If the
connections are ok, it is not possible to test this CT due to its too high
inductance.
111.xxx
Warning [111.xxx] Inductance of load too low.
Reason:
Measurement frequency is too high and cannot be decreased because of too
low CT inductance.
Solution:
The test cannot be continued. Check the connections to the CT. If the
connections are ok, it is not possible to test this CT due to its too low
inductance.
112.xxx
Warning [112.xxx] Measurement error, reduce noise level.
Reason:
It is not possible to obtain stable measurement results due to incorrect wiring,
wiring short-circuit, connected load or external interferences.
Solution:
Check the wiring. The primary side of the CT must not be short-circuited.
Disconnect the primary side of the CT from the transmission lines, connect one
primary side terminal to PE and open the other one. Use the delivered original
coax cables for measurement (see section 2.7.2 ”Noise Reduction
Techniques” on page 43). Make sure that no burden is connected to the
secondary side of the CT. This error may also indicate a winding short circuit
of the CT.
Содержание CT Analyzer
Страница 1: ...CT Analyzer CT SB2 User Manual...
Страница 3: ...CT Analyzer User Manual...
Страница 4: ...CT Analyzer User Manual 2 Manual Version CTAnalyzerUM AE 3...
Страница 39: ...37 Setup and Connection Figure 2 15 Measurement setup for delta compensation Ratio 1 L2 L3 L1...
Страница 50: ...CT Analyzer User Manual 48...
Страница 60: ...CT Analyzer User Manual 58...
Страница 80: ...CT Analyzer User Manual 78...
Страница 162: ...CT Analyzer User Manual 160...
Страница 172: ...CT Analyzer User Manual 170...
Страница 188: ...CT Analyzer User Manual 186...
Страница 194: ...CT Analyzer User Manual 192...
Страница 195: ...CT SB2 User Manual...
Страница 196: ...CT SB2 User Manual 2 Manual Version CTSB2 AE 3...
Страница 204: ...CT SB2 User Manual 10...
Страница 206: ...CT SB2 User Manual 12...
Страница 211: ...17 Hardware 1 x Cable bag Bag for cables and accessories Accessories Description...
Страница 212: ...CT SB2 User Manual 18...
Страница 226: ...CT SB2 User Manual 32...
Страница 242: ...CT SB2 User Manual 48...
Страница 248: ...CT SB2 User Manual 54...
Страница 250: ...CT SB2 User Manual 56...
Страница 252: ...CT SB2 User Manual 58...