8.4.3 Re-performing a calibration
A calibration can be performed by clicking on the corresponding calibration icon in the Home ribbon.
•
Clicking on a calibration icon (e.g. Full-Range Gain calibration
) opens the calibration
dialog
•
The calibration measurement can be started by clicking the start button
.
•
After the calibration is
the calibration dialog can be closed
•
The calibration icon turns green indicating that the calibration is now active
.
From now on, this calibration is used for the measurement.
A calibration can be disabled and enabled by clicking on its calibration icon.
To re-performed a calibration, open the calibration dialog by clicking on the arrow on the
bottom of the calibration icon
. Then select
.
Calibration can then be re-performed by pressing the
Start
button.
8.4.4 Full-Range calibration below 10 Hz
Full-Range calibration measures between 10 Hz and 50 MHz at factory pre-defined frequencies. Full-
Range calibration normally starts at 10 Hz to reduce calibration time.
If you decide that frequencies between 1 Hz and 10 Hz must be included in the Full-Range
calibration, you must set the
Start Frequency
to a value <10 Hz before executing the Full-
Range calibration.
Alternatively User-Range calibration might be more applicable in that case.
Full-Range calibration frequency range:
•
If Start Frequency is set ≥10 Hz, then Full-Range calibration will run from 10 Hz to 50 MHz
•
If Start Frequency is set <10 Hz, then Full-Range calibration will run from 1 Hz to 50 MHz
If a Full-Range calibration has been performed at a
Start Frequency
>10 Hz and
Start
Frequency
is changed to a value <10 Hz after calibration, the correction value at 10 Hz
will be extrapolated to 1 Hz. This is indicated by an
orange
Full-Range calibration icon.
Re-performing calibration will run the calibration from 1 Hz and the icon will turn green
again.
Note that the calibration takes significantly longer when it starts at 1 Hz.
90
OMICRON Lab
Содержание Bode 100
Страница 1: ...Smart Measurement Solutions Bode 100 User Manual...
Страница 24: ...Figure 6 3 Collapsing the measurement and trace configuration areas 24 OMICRON Lab...
Страница 45: ...Figure 7 3 Context menu of chart Measurement types and applications OMICRON Lab 45...
Страница 83: ...Open B WIC Open B SMC Short B WIC Short B SMC Load B WIC Load B SMC Calibration OMICRON Lab 83...