8.3.1 Calibrating a Reflection or One-Port Impedance measurement
In this section you learn how to calibrate an Impedance, Reflection or Admittance measurement in the
Transmission / Reflection measurement mode
or in the
One-Port impedance measurement
mode
.
The Transmission / Reflection mode offers the possibility to measure Transmission and Reflection
sequentially. Please note that the Impedance (Open/Short/Load) calibration is only applied to a
Impedance, Reflection or Admittance measurement. If you want to calibrate a Gain measurement
please refer to
8.2.1 Calibrating a Transmission (S21) measurement
In the
Transmission / Reflection measurement mode
or in the
One-Port impedance measurement
mode
both receivers are internally connected to the 50 Ω source resistance.
Bode 100
is internally
calibrated such that it measures the Impedance/Reflection directly at the OUTPUT port of the device.
The reference plane is directly at the OUTPUT port (BNC connector).
When connecting a DUT with a coaxial cable, the cable introduces additional impedance and time-
delay. To compensate these effects, the reference plane can be moved to the end of the cable by
performing an Open/Short/Load calibration at the end of the cable.
Figure 8-4: Shifting the reference plane by performing Open, Short and Load Calibration
78
OMICRON Lab
Содержание Bode 100
Страница 1: ...Smart Measurement Solutions Bode 100 User Manual...
Страница 24: ...Figure 6 3 Collapsing the measurement and trace configuration areas 24 OMICRON Lab...
Страница 45: ...Figure 7 3 Context menu of chart Measurement types and applications OMICRON Lab 45...
Страница 83: ...Open B WIC Open B SMC Short B WIC Short B SMC Load B WIC Load B SMC Calibration OMICRON Lab 83...