7.4.2 Impedance Adapter
The impedance adapter measurement mode is especially designed for measurements with the
OMICRON Lab impedance test fixtures
B-WIC
and
B-SMC
. The impedance adapter measurement
mode ensures that
Bode 100
is configured correctly to achieve best results when measuring with both
B-WIC
and
B-SMC
.
Calibration
The impedance adapter measurement mode requires impedance (open, short, load) calibration.
For more details on how to calibrate the
B-WIC
or
B-SMC
impedance adapter, please refer to
8.3.3 Calibrating an Impedance Adapter measurement
Measurement information
With
B-WIC
and
B-SMC
, the dynamic range of both input channels is used. This widens the usable
impedance measurement range to 20 mΩ to 600 kΩ.
It is recommended to use the 0.5 m BNC cables delivered with
Bode 100
to connect
B-WIC
or
B-SMC
to the
Bode 100
.
B-WIC
and
B-SMC
are designed to measure physically small DUTs. Stray-fields between the DUT and
the grounded housing of the impedance adapters might introduce a systematic measurement error.
The error is negligible when measuring physically small objects. To measure physically big DUTs it is
recommended to use a grounded measurement configuration such as shown in
Typical applications
Typical applications for the impedance adapter measurement mode are:
•
Measuring impedance of capacitors, inductors and other passives (THT and SMC).
•
Measuring impedance piezo-elements or quartz elements.
Impedance Adapter measurement example
In this measurement example we will determine the inductance (Ls) and the equivalent series
resistance (Rs) of an inductance with the
B-WIC
impedance measurement adapter for through-hole-
type components. Connect the
B-WIC
to
Bode 100
using three BNC cables as shown below.
Measurement types and applications
OMICRON Lab
59
Содержание Bode 100
Страница 1: ...Smart Measurement Solutions Bode 100 User Manual...
Страница 24: ...Figure 6 3 Collapsing the measurement and trace configuration areas 24 OMICRON Lab...
Страница 45: ...Figure 7 3 Context menu of chart Measurement types and applications OMICRON Lab 45...
Страница 83: ...Open B WIC Open B SMC Short B WIC Short B SMC Load B WIC Load B SMC Calibration OMICRON Lab 83...