4.2 Bode 100 specifications
In this section you can find the most important technical data for the
Bode 100
. This
technical data is valid for
Bode 100 R2
(Hardware revision 2). To find out what hardware
revision you use, please check
Technical data can change without notice. You can download a detailed technical data sheet
from the OMICRON Lab website
→ Bode 100 → Technical Data
Table 4-2:
Bode 100
specifications
Characteristic
Rating
OUTPUT
Voltage range
-30 dBm...13 dBm
7 mVrms...1 Vrms at 50 Ω load
14 mVrms...2 Vrms at high impedance load
Frequency range
1 Hz...50 MHz
Wave form
sinusoidal
Source impedance
50 Ω
Connector
BNC
INPUT CH 1, INPUT CH 2
AC measurement range
100 mVrms full scale with 0 dB input attenuation
10 Vrms full scale with 40 dB input attenuation
Frequency range
1 Hz...50 MHz
Input impedance
Selectable via software:
•
50 Ω
•
1 MΩ ±2% || 40...55 pF
Connectors
BNC
Input attenuators
Selectable via software:
0 dB, 10 dB, 20 dB, 30 dB, 40 dB
Receiver bandwidth
Selectable via software:
1 Hz, 3 Hz, 10 Hz, 30 Hz, 100 Hz, 300 Hz, 1 kHz,
3 kHz , 5 kHz*
*only
Bode 100
R2
Dynamic Range
> 100 dB (10 Hz receiver bandwidth)
Gain Error
< 0.1 dB (calibrated)
Phase Error
< 0.5 ° (calibrated)
PC interface
Universal serial bus
USB 2.0 or higher
Connector
USB type B socket
Technical data
OMICRON Lab
11
Содержание Bode 100
Страница 1: ...Smart Measurement Solutions Bode 100 User Manual...
Страница 24: ...Figure 6 3 Collapsing the measurement and trace configuration areas 24 OMICRON Lab...
Страница 45: ...Figure 7 3 Context menu of chart Measurement types and applications OMICRON Lab 45...
Страница 83: ...Open B WIC Open B SMC Short B WIC Short B SMC Load B WIC Load B SMC Calibration OMICRON Lab 83...