10 |
P a g e
4.4
Thermocouple Configuration
The SP-016 uses a thermocouple input for temperature compensation and provides interfaces to type J, K, T, E, N, R, S, B,
and C thermocouples with the capability of enabling or disabling the open detect feature. To use these features, follow
these steps:
Step 1:
Click the Inputs configuration tab on SYNC and choose an input type that includes TC from the Type dropdown.
Step 2:
Click the input you wish to configure and select your thermocouple type from the Device Range/Type dropdown.
Step 3:
Click the Open Detect drop down and choose to enable or disable it.
Type
Range
Accuracy
J
-210°C and 1200°C
0.4°C
K
160°C to 1372°C
0.4°C
T
190°C to 400°C
0.4°C
E
-220°C to 1000°C
0.4°C
N
-100°C to 1300°C
0.4°C
R
40°C to 1768°C
0.5°C
S
100°C to 1768°C
0.5°C
B
640°C to 1820°C
0.5°C
C
°0C to 2320°C
0.4°C