25 |
M 5 7 7 9
6.3.4.2
Sensor Trigger Function
The Sensor Trigger function is used to reset the IPSO min/max values and to control the Calibration
process.
Sensor Trigger Function
7
6
5
4
3
2
1
0
Reset
Min/Max
0
0
0
0
0
0
0
?
15
14
13
12
11
10
9
8
0
0
0
0
0
0
0
0
Setting the Reset Min/Max bit to 1 will reset the Min/Max values recorded by the IPSO process. No
user calibration is supported in the temperature sensor and all configuration bits should be written
as 0.
6.4
Discrete I/O Interface
The SP-006 supports a DIO Interface that provides 2 discrete inputs that are hardwired to the outputs. These may be
used to detect the state of external switches (output off) or to monitor the state of the outputs.
6.4.1
DIO Descriptor
Offset
Name
Value
Description
0x00
Sensor Type
0x18
Digital Type (Bit mapped)
0x01
Data Type/Format
0x46
Configurable, Float type
0x02
Configuration
0x23
Scaling applied, Bits 0 and 1 enabled
0x03
Sensor Device
0x0f
DIN bits enabled / inverted
0x04
UOMR
“DIN”
Units of measure
6.4.1.1
DIO Sensor Type
The interface provides a bitmapped input of the 2 digital signal lines.
Sensor
Type
SI Derived
Units
Measurement
0x18
DIN
Bitmapped digital inputs
6.4.1.2
DIO Data Type/Format
DIO Data Type/Format
7
6
5
4
3
2
1
0
Smart
Sensor
Writeable
Factory
Calibrate reserved
Data Type
0
0
0
0
6 == Floating point
6.4.1.2.1
Data Type
The 4-bit Data Type field determines the type of data of the specific sensor.