OS301A-USB Series

OS150A-2USB Series

Infrared Temperature Sensors

e-mail: [email protected]

For latest product manuals:

www.omegamanual.info

Shop online at 

omega.com

User’s Guide

TM

Содержание OS301A-USB Series

Страница 1: ...OS301A USB Series OS150A 2USB Series Infrared Temperature Sensors e mail info omega com For latest product manuals www omegamanual info Shop online at omega com User sGuide TM...

Страница 2: ...rves the right to alter specifications without notice Servicing North America U S A Omega Engineering Inc Headquarters Toll Free 1 800 826 6342 USA Canada only Customer Service 1 800 622 2378 USA Cana...

Страница 3: ...als such as many reflective metals They are also capable of measuring through glass viewports GENERAL SPECIFICATIONS Series OS150A 2USB OS301A USB Spectral Response 2 2 m 8 to 14 m Application Ferrous...

Страница 4: ...nmental Rating IP65 Ambient Operating Temperature 0 C to 70 C cooled models are available for higher temperatures Relative Humidity 95 max non condensing DIMENSIONS MINIMUM MEASURABLE TEMPERATURE mode...

Страница 5: ...MODEL NUMBERS...

Страница 6: ...m Spot Dia inches Spot Dia mm Distance Sensor to object inches Distance Sensor to object mm 15 100 215 0 6 3 9 8 5 0 59 1 118 1 59 1 118 1 0 1500 3000 0 1500 3000 D S 25 1 D S 15 1 OS251A 2USB Distanc...

Страница 7: ...ocation where energy from lamps heaters and sunlight cannot be reflected from the target into the lens The use of shields may help in this respect For further information and assistance contact Omega...

Страница 8: ...and adjustable mounting bracket accessories which are shown below AIR WATER COOLED HOUSING The air water cooled housing shown below allows the sensor to withstand high ambient temperatures It is equi...

Страница 9: ...e 4 20 mA output connect the sensor to the controller indicator with the 24 V DC power supply in the loop using the following table and connection diagram as a guide The maximum loop impedance is 900...

Страница 10: ...System Requirements Windows Vista or newer USB 2 0 port Internet access for software download Installation 1 Download the software from the Omega website 2 Unzip and run setup exe 3 Follow the on scr...

Страница 11: ...ng 0 Peak valley hold processing If required hold processing can be applied by setting Hold Mode to Peak or Valley and setting the hold period This is useful if the temperature reading is interrupted...

Страница 12: ...1 Temperature range OS150A 2USB only 0 LT 1 PT 2 MT 3 HT R 0x05 1 Modbus slave address R W 0x06 1 Reflected Energy Compensation 0 Off 1 On R W 0x07 1 Reflected Temperature R W 0x08 1 Emissivity 1 LSB...

Страница 13: ...the exception of addresses 0x17 and 0x18 all write operations are saved to non volatile memory 3 For further information please refer to http www modbus org specs php 4 The default Modbus slave addre...

Страница 14: ...dvice LENS CLEANING Keep the lens clean at all times Any foreign matter on the lens would affect measurement accuracy Blow off loose particles if not using the air purge accessory with an air puffer T...

Страница 15: ...or otherwise shall not exceed the purchase price of the component upon which liability is based In no event shall OMEGA be liable for consequential incidental or special damages CONDITIONS Equipment s...

Страница 16: ...ne Paddlewheel Systems M U Totalizers Batch Controllers pH CONDUCTIVITY M U pH Electrodes Testers Accessories M U Benchtop Laboratory Meters M U Controllers Calibrators Simulators Pumps M U Industrial...