M20013 REL-0000 (2103)
User Manual HHP460 Smart Manometer with Altimeter & Air Speed Test
47 of 72
User Calibration button
The
User Calibration
button displays options to calibrate
the sensor.
The
User Calibration
button displays a
yellow alert icon
to indicate you have changed a setting and you have not
yet saved the setting to the Smart Manometer.
User Calibration: For authorized personnel only
What does “Calibration Enabled” refer to?
▪
With a check mark in the box, Calibration Enabled displays the
Live Reading
with your current user calibration applied.
▪
Without a check mark in the box, it displays
Live Reading
without your user calibration.
Restore Factory Defaults
This button clears your user calibration from the sensor.
Choose Calibration Unit
▪
Select the calibration reference units.
▪
This does not affect or depend on the configuration units.
Begin button…Apply button
1. Click
Begin
.
2. Select the point.
3. Apply reference pressure within the range shown.
4. Enter the reference value.
5. Save the point.
6. Repeat steps 2-5 until you are done. (You have to change at
least one point.)
7. Click the
Apply
button.