M20013 REL-0000 (2103)
User Manual HHP460 Smart Manometer with Altimeter & Air Speed Test
68 of 72
EMC compliance
Conformity to EN 61326-1:2016
This product has been evaluated to EN 61326-1:2016 the EMC
General requirements for electrical equipment for measurement,
control and laboratory use. This evaluation demonstrated that
when the product is exposed to Radiated Radio Frequency
energy in the frequency range 130-1000 MHz and 1.6-2.1 GHz, the
pressure and temperature readings may fluctuate beyond the
published tolerance of each respective value being measured. In
the event of such a phenomenon occurring, the operator should
move away from (or have removed) any source of strong RF
emissions such as radio transmitters, cell-phones, etc., and repeat
the measurement, or reconnect to a PC.
Electrostatic discharge (ESD)
This unit is sensitive to ESD discharges that may result in the
unit re-starting of the operating system, or interruption of a PC
connection. In the unlikely event of such an occurrence, users
should ensure they discharge any static charge they may be
carrying by touching a known grounded conductive surface prior
to operating the equipment .