Instrument Description
1-1
1. I
NSTRUMENT
D
ESCRIPTION
Specifications
G
ENERAL
S
PECIFICATIONS
:
Basic Accuracy
±(0.04% |rdg| + 0.3 °C)
1
Conformity
ITS-90
Temperature
Ranges
°C
°F
K
K
-200 to 1372
-328 to 2502
73 to 1645
J
-210 to 1200
-346 to 2192
63 to 1473
T
-250 to 400
-418 to 752
23 to 673
E
-250 to 1000
-418 to 1832
23 to 1273
B
600 to 1820
1112 to 3308
873 to 2093
N
-200 to 1300
-328 to 2372
73 to 1573
R
0 to 1760
32 to 3200
273 to 2033
S
0 to 1760
32 to 3200
273 to 2033
Connector Type
One (1) Mini-TC (HH931T)
Two (2) Mini-TC (HH932T)
Probe Zero
Function
Resolution 0.1 °C/°F/K
Display
Four (4) digit LCD, with Temperature, Units, Function, Trend,
Polarity, Battery, Data, Real-time Clock, and Decimal Indicators
Display Backlight
Four (4) LED Backlight with 30-second timeout
Display Resolution
0.1
(-99.9 to 999.9 °)
1 ° (< -99.9 °, > 999.9 °)
Reading Rate
3 / Second for Readings and Trend Indicators
Battery Type
3 AA (IEC LR6, ANSI 15) Alkaline
Battery Life
1000 Hours Typical
Battery Indicator
Four (4) Stage Battery Charge Indicator
Statistics
Minimum Reading
Maximum Reading
Average Reading
Reading Range
Standard Deviation
T1–T2 (HH932T Only)
Keypad
Twelve (12) momentary switches with audible and tactile
feedback
Clock
Real-Time Clock and Elapsed Statistics Run Time
Power Cycle
Configuration
Retention
Instrument retains:
-
Sensor Type
-
Temperature Unit
-
Offset Values
-
Saved Measurements
-
Real-Time Clock
Input Current
±50 nA
Содержание HH931T
Страница 39: ...Appendices B i B EXPANDED INSTRUMENT UNCERTAINTIES Thermocouple Type B ...
Страница 40: ...Appendices B ii Thermocouple Type E ...
Страница 41: ...Appendices B iii Thermocouple Type J ...
Страница 42: ...Appendices B iv Thermocouple Type K ...
Страница 43: ...Appendices B v Thermocouple Type N ...
Страница 44: ...Appendices B vi Thermocouple Type R ...
Страница 45: ...Appendices B vii Thermocouple Type S ...
Страница 46: ...Appendices B viii Thermocouple Type T ...