background image

M2218/1201

http://www.omega.com  

e-mail: [email protected]

®

User’s Guide

CCT-80

Strain Gage Signal Conditioners

Содержание CCT-80 Series

Страница 1: ...M2218 1201 http www omega com e mail info omega com User sGuide CCT 80 Strain Gage Signal Conditioners...

Страница 2: ...France 0800 4 06342 e mail france omega com Germany Austria Daimlerstrasse 26 D 75392 Deckenpfronn Germany Tel 49 07056 3017 FAX 49 07056 8540 Toll Free in Germany 0130 11 21 66 e mail germany omega...

Страница 3: ...Changing the Power Supply 4 Signal Output module MS 5 Changing the Signal Output 5 Signal Input module ME Overview 6 CCT 80 Electrical Features Specifications 7 Signal Input Range Selection 8 Adjustm...

Страница 4: ...the case the following must be checked The front label blue color must be in its correct position with terminals 1 and 7 power supply separated from the other terminals The three internal modules must...

Страница 5: ...e Transmitter to control a machine or process where injury to personnel or damage to equipment or process may occur as a result of failure of the Transmitter Power Fuse supply value 230 Vac 50 mA 115...

Страница 6: ...4 to 20 mA unless specified otherwise To select a 0 to 10 V output remove jumpers E and F as shown in Figure 6 Other non standard output voltage and current ranges may be obtained by adding and or re...

Страница 7: ...if closed Adjust the low range level of the output Offset negative fine AMPLIFIER GAIN Jumper B if closed Gain at maximum level Jumper C if closed Gain at medium level Jumpers B and C opened Gain at...

Страница 8: ...SUMPTION 1 5 VA TEST VOLTAGE 4 kVeff 50 Hz 1m This signal conditioner provides connection to 4 or 6 wire strain gages load cells which provide a nominal of 2 mV V or 3 mV V output with 350 impedance S...

Страница 9: ...als to be used 5 Power up the signal conditioner with the correct power supply 6 Adjust the calibrator until it generates the low signal level 0 Vdc 7 Turn the ZERO trimmer P1 located on the ME input...

Страница 10: ...ENTACION TARE ADJUST When internal tare adjust on the ME input board is not used the signal conditioner terminals 3 and 4 must not be connected POWER SUPPLY SIGNALOUTPUT Current SIGNALOUTPUT Voltage O...

Страница 11: ...Lo H i TECHNICAL DATA WEIGHT 270 g HOUSING BASE Polycarbonate RAL 7032 UL 94 V 1 light grey IP 40 TERMINAL HOUSING COVER AND BLIND PLUGS Polycarbonate UL 94 V 2 dark grey IP 20 WIRE CROSS SECTION 4 mm...

Страница 12: ...NOTES 11...

Страница 13: ...rwise shall not exceed the purchase price of the component upon which liability is based In no event shall OMEGA be liable for consequential incidental or special damages CONDITIONS Equipment sold by...

Страница 14: ...ystems M U Totalizers Batch Controllers pH CONDUCTIVITY M U pH Electrodes Testers Accessories M U Benchtop Laboratory Meters M U Controllers Calibrators Simulators Pumps M U Industrial pH Conductivity...