2
Settings
Subdisplay value: 000.0 to 999.9
Main display value: 0000 to 9999
Main display decimal place: XXXX, X.XXX, XX.XX, or XXX.X
Main display unit: A, mA, μA, °C, MΩ, kΩ, Ω, kHz, Hz, μF, nF, %, lx, no unit, V, or mV
Example
For a current clamp sensor with an input-to-output ratio of 0 mV/A,
set the subdisplay to 0.0 mV, and set the main display (value, decimal
place, and unit) to 0.00 A.
With these settings, a maximum input of 660.0 mV will be displayed as
66.00 A.
0
1 A
10 mV
66 A
660 mV
Display
Input
4.6.3 Current Span Setting for the 100 mA Range for DCmA Measurement
When you select the 00 mA range in DCmA measurement mode (
), you can change the
current span (0 to 00 mA, 0 to 0 mA, or 0 to 0 mA).
Input Current Value
(Main display)
Subdisplay for the 00 mA Range
0 to 00 mA
0 to 0 mA
0 to 0 mA
-0.00 mA
-110.0%
-300.0%
-220.0%
0.00 mA
0.0%
-25.0%
0.0%
0.00 mA
10.0%
0.0%
20.0%
0.00 mA
50.0%
100.0%
100.0%
00.00 mA
100.0%
225.0%
200.0%
0.00 mA
110.0%
250.0%
220.0%
1) Press the LIGHT or SHIFT key to display “SPAn.”
HOLD
(The selected current span appears in the subdisplay. The default span is 0 to 00 mA.)
2) Use ▲ (RANGE) and ▼ (REL) to select 0 to 100 mA, 10 to 50 mA, or 0 to 50 mA.
3) Press the HOLD key to save the setting. (After “SEt” appears, “SPAn” will appear
again.)
) To return to normal measurement from setup mode, hold down the LIGHT key for
second or longer.