4.4.1 Constant Current Output (SOURCE mode)
) Turn the function switch to the “
” position.
After you turn the function switch, “OUTPUT” appears on the screen, and the output is
set to 0% of the current span setting.
(If you turn the function switch from the “
” (current sweep) position to the “
”
position, the output value setting and SOURCE/SIMULATE setting are retained.)
2) If “SIMULATE” is displayed, press the SHIFT key.
(“SIMULATE” disappears, the CA0 switches to SOURCE mode, and current output
starts.)
3) Connect the lead cables to the positive (+) and negative (–) SOURCE terminals.
) Use the output value setting keys to set the output value.
Black
Red
Lead cables
Output Value
Setting Key
Output Setting
(Increase and decrease)
%STEP
*
▲ 1-step (25%) increase
▼ 1-step (25%) decrease
COARSE
▲ 0. mA increase
▼ 0. mA decrease
FINE
▲ 0.00 mA increase
▼ 0.00 mA decrease
* For details about using the %STEP key, see “Output
Values for Each Step” and “Span Check Mode.”
) Connect the lead cables to the circuit under test.
Tip
• Because the CA0 outputs current in SOURCE mode, it consumes more
current than in other modes. You can keep the current consumption down in
SIMULATE mode.
• Hold down the %STEP, COARSE, or FINE key to increase or decrease
each digit consecutively.
Output Values for Each Step
You can press %STEP▲ or %STEP▼ to increase or decrease the output by one step (25%).
You can check the step value (as a percentage) in the subdisplay. (The output current
appears in the main display.)