Quick Reference Guide to Near Infra-red Imaging with VISERA ELITE II

Disclaimer:

 Information provided in this guide is not exhaustive. Please refer to the complete Instructions For Use for each 

item of equipment which should be thoroughly read and understood prior to handling, use and storage of the equipment. 

Light Source

Camera Head

Switch On

Switch on the camera system using the workstation master 
switch which will then turn from white to green.

There is a secondary power switch on the transformer 
under the workstation. This should remain switched on.

Insert a NIR compatible camera head (CH-S200-XZ-EB 
or CH-S200-XZ-EA) into the camera processor, ensuring 
the connector is the correct way up.

UP mark

Video connector

Electrical contacts

Video connector socket

UP mark

Video connector

The colour bars on the monitor will now be replaced with 
an image from the camera head.

Insert the light guide cable into the separate NIR light 
source until it clicks.

Warning:

 When turning ON the light source, never allow 

the distal end of an endoscope or a light guide cable to 
come into contact with the patient and/or other flammable 
materials, such as operating room drapes. Patient injury 
and/or a fire can result.

Remove any light guide cable or ENDOEYE from the output 
socket of the processor.

Output socket

Video connector socket

OTV

Output socket

 

Issue 1 

05/02/2021

 

KMF-08928 

NCDR 02866