CMOS color camera SC180

20

6

Troubleshooting

The following text describes typical errors that may occur when using the camera. 

6.1

 Contacting Customer Service

Please contact your local OLYMPUS support if you have any questions about the
product. The addresses can be found on the following websites:

Life Science Solutions

https://www.olympus-lifescience.com/en/contact-us/

Industrial Solutions

https://www.olympus-ims.com/en/contact-us/

Error

Possible causes

Solution

The camera can't be 

found

The camera was not 

selected during the installa-

tion of your software.

The camera driver has to be installed. To do so, start the 

installation wizard for the software again.

Browse to the directory that contains the cellSens or 

OLYMPUS Stream software and double click the 

setup.exe

 file. The installation wizard starts. 

In the next dialog boxes, select the options 

Install or 

maintain imaging software

 > 

Maintain imaging software

 > 

Provide licenses, update installation

.

When the 

Image source selection

 dialog box is 

displayed, select the 

Olympus

 manufacturer and select 

the check box next to the camera you want to use for 

image acquisition. 

Follow the further instructions of the installation wizard 

and finish the installation.

A faulty image appears

A power saving mode in the 

BIOS prevents image 

acquisition.

If possible, ask your system administrator to switch off all 

power saving modes in the BIOS of your PC. 

Please note: 

Changes in the BIOS (Basic Input Output System) can 

have a negative effect on the way your computer oper-

ates. Make sure that only qualified personnel make 

changes to the settings in the BIOS.

When starting the live-

image, the “Camera time-

out “error message 

appears.

The USB cable you are 

using is too long.

Use the supplied USB cable.

Other processes that 

require a lot of processing 

power are running at the 

same time.

Check which other processes are running and end those 

that are not required.

Содержание SC180

Страница 1: ...ish Optical Microscope Accessory To obtain optimum performance of this camera and to ensure the safety study this manual thoroughly before operating the camera and keep it on hand during operation of...

Страница 2: ...damages arising out of purchase or use of this manual or the information contained herein No part of this document may be reproduced or transmitted in any form or by any means electronic or mechanical...

Страница 3: ...ymbols used 7 2 2 Avoiding personal injuries 8 2 3 Avoiding device damage 9 3 System environment 10 3 1 General remarks 10 3 2 Microscopes and camera adapters 12 3 3 PC 13 3 4 Software 14 4 Installati...

Страница 4: ...essential components To be able to use the camera you ll also need the following components Computer with MS Windows operating system Light microscope with camera adapter or something similar for exam...

Страница 5: ...ng time but also in a lower resolution Technical data SC180 Chip type CMOS Chip size 1 2 3 inch Effective area width x height 6 140 mm x 4 605 mm Pixel size 1 25 m x 1 25 m Maximum resolution 4912 x 3...

Страница 6: ...s The maximum permissible range for the ambient temperature during camera operation is 5 C to 40 C The maximum permissible relative humidity during camera operation is 80 condensation free When the ca...

Страница 7: ...re and during the installation 2 1 Explanation of the symbols used Personal injuries Device damage General notes The warning sign and the word WARNING indicate dangerous situations that can lead to se...

Страница 8: ...the case of a fire the material of the camera can give off toxic gases When these are inhaled it can cause irritation of the respiratory passages and damage to them Wear respiratory protection when fi...

Страница 9: ...ensitive and must not come into contact with bare hands or any other objects otherwise the acquisitions will be impaired by fingerprints or scratches Remove dust from the protective glass carefully wi...

Страница 10: ...e used in combination with many available microscopes However its performance in combination with non Olympus microscopes has not been tested Application context Explanation Countermeasure Using objec...

Страница 11: ...the auto exposure to fail resulting in an underexposed image In the Camera Control of the software define an exposure compensation Alternatively switch to manual exposure mode Acquiring images with a...

Страница 12: ...not listed there Upright microscopes BX61 BX51 BX41 CX41 CX31 MX61 CX23 CX33 CX43 MX61L MX51 BX43 BX46 BX53 BX63 Inverted microscopes IX81 IX71 IX51 IX83 IX73 IX53 CKX53 Stereo microscope SZX16 SZX10...

Страница 13: ...mera with a laptop Generally speaking it is possible to operate the camera with a laptop If the power supply is not sufficient to operate the camera when an integrated USB interface is used connect th...

Страница 14: ...ized software interface between software programs and image acquisition hardware such as digital cameras or scanners You can download the DP2 TWAIN software from this Olympus website https www olympus...

Страница 15: ...are using when installing the software The correct camera driver will then be automatically installed Do the following 1 Close all running application programs including the application programs that...

Страница 16: ...on dialog box select the Olympus manufacturer Select the SC180 check box 5 At the end of the installation the files are copied to the PC You can then finalize the installation procedure The software a...

Страница 17: ...uaranteed when this cable is used 4 Plug the end of the camera s USB cable with the type A plug into a free USB 3 0 port of your PC If possible use a USB 3 0 port that is located on the back side of y...

Страница 18: ...CMOS color camera SC180 18 The camera has now been registered as a device and is ready for use Now the camera driver appears in the Device Manager 6 Start the software that has already been installed...

Страница 19: ...an gauze moistened with pure alcohol Cleaning the housing Parts other than the glass components should be cleaned by wiping with a clean cloth Do not use organic solvents to remove major stains Use a...

Страница 20: ...ns Install or maintain imaging software Maintain imaging software Provide licenses update installation When the Image source selection dialog box is displayed select the Olympus manufacturer and selec...

Страница 21: ...h the European Union directive on the restriction of the use of certain hazardous substances in electrical and electronic equipment 2011 65 EU 7 4 FCC conformity USA This device complies with part 15...

Страница 22: ...ent or relocate the receiving antenna 2 Increase the separation between the equipment and receiver 3 Connect the equipment into an outlet on a circuit different from that to which the receiver is conn...

Страница 23: ...7 Notes on conformity and disposal 23...

Страница 24: ...63 0914 Japan OLYMPUSEUROPASE CO KG Wendenstrasse 14 18 20097 Hamburg Germany OLYMPUSCORPORATIONOFTHEAMERICAS 3500 Corporate Parkway P O Box 610 Center Valley PA 18034 0610 U S A OLYMPUSCORPORATIONOFA...

Отзывы: