Using Software Options
Part # 910-250C
89
When TVG mode is active, any parameter that would affect instrument calibration is
automatically locked. This includes: [GAIN], [ZERO OFFSET], [VELOCITY],
[ANGLE], [REJECT], and all pulser and filter parameters.
Once the TVG setup is complete, you can activate the Reference Gain function by
pressing [2
nd
F], [GAIN] (REF DB). This locks in a reference gain level and allows
scanning gain to be added or subtracted to accommodate specific applications. Note that
the additional scanning gain cannot be added to the original reference gain level as this
would change the calibration and invalidate the TVG setup.
9.3.2
Saving and Recalling TVG Setups
TVG setups are saved and recalled in the same manner as any general waveform/
calibration. With TVG mode active, the [SAVE WAVE] key is used to save the entire
setup in the datalogger. When a setup that was stored with a TVG setup is recalled, the
instrument automatically comes up in TVG mode.
9.3.3
Exiting TVG Mode
To clear the TVG setup and return to the normal operation, follow these steps:
1. Press [OPTION].
2. Use the slewing keys to highlight the TVG selection and press [ENTER] to go into
the TVG setup.
3. Use the slewing keys to highlight Off and press [ENTER].
4. Press [OPTION] to return to the A-Scan display.
9.4
Advanced DAC/TVG
9.4.1
Description
The Advanced DAC/TVG (ADT) software option for the EPOCH 4PLUS significantly
increases the functionality of the standard DAC/TVG software, allowing users to
customize DAC/TVG setups to their unique application requirements. The ADT feature
incorporates capabilities that adhere to ASME, ASME-3 and JIS sizing codes of Olympus
NDT
TM
standard DAC/TVG software with direct control of gain, range, and zero offset, as
well as reference and transfer correction. In addition, the ADT option provides new
features such as 20%-80% DAC curves, customizable DAC curves, and a user-defined
TVG table to meet advanced and unique inspection needs. ADT also allows the user to
toggle between DAC and TVG views in many of its functions, allowing the operator to use
both techniques during a single inspection.
Note:
For a definition and detailed description of a general Distance Amplitude
Correction (DAC) curve and Time Varied Gain (TVG), see manual sections
9.2 and 9.3.
Содержание Panametrics Epoch 4 Plus
Страница 4: ...EPOCH 4PLUS...
Страница 10: ...EPOCH 4Plus...
Страница 18: ...8 EPOCH 4PLUS...
Страница 40: ...30 EPOCH 4PLUS...
Страница 50: ...40 EPOCH 4PLUS...
Страница 54: ...44 EPOCH 4PLUS...
Страница 92: ...82 EPOCH 4PLUS...
Страница 172: ...162 EPOCH 4PLUS...
Страница 184: ...174 EPOCH 4PLUS Figure D 3 VGA Adapter Cable Olympus NDT Part EP4 C 15VGA 6 Figure D 4 VGA Adapter Cable Wire Run List...