DMTA-10040-01EN, Rev. E, February 2018
Control Functions
87
Figure 4
‑
12 The EDDY CURRENT SLIDE RULE menu
4.3.4
Display Menu — DISP Key
The display menu contains submenus to control various functions such as horizontal
and vertical position, display erase, trace, grid, and zoom.
DSP MODE (display mode)
Six display modes are available on the NORTEC 600 instrument:
IMP
(impedance),
SWP+IMP
(sweep plus impedance),
SWEEP
,
WATERFALL
,
DUAL
IMP
(dual impedance), and
ALL IN ONE
.
To adjust the display mode, press the DISP menu key (
), followed by the
A key. With
DSP MODE
highlighted, rotate the knob until the desired display
mode is displayed.
IMP
Impedance mode is the most common display mode. It incorporates a
10 × 10
grid configuration on the screen. The eddy current signal is shown moving
horizontally and vertically.
SWEEP
Commonly used with rotating scanners. The eddy current signal is shown
moving horizontally at a fixed rate across the screen.
SWP+IMP
(NORTEC 600S and NORTEC 600D models only)
Typically used while setting up an inspection using a rotating scanner.
Impedance (
IMP
) and
SWEEP
screens are displayed together on a “split
screen.”
Содержание nortec 600
Страница 8: ...DMTA 10040 01EN Rev E February 2018 Table of Contents viii...
Страница 16: ...DMTA 10040 01EN Rev E February 2018 Labels and Symbols 6...
Страница 30: ...DMTA 10040 01EN Rev E February 2018 Introduction 20...
Страница 58: ...DMTA 10040 01EN Rev E February 2018 Chapter 1 48...
Страница 71: ...DMTA 10040 01EN Rev E February 2018 Software User Interface 61 Press the Return key to exit...
Страница 72: ...DMTA 10040 01EN Rev E February 2018 Chapter 2 62...
Страница 342: ...DMTA 10040 01EN Rev E February 2018 Chapter 7 332...
Страница 356: ...DMTA 10040 01EN Rev E February 2018 Appendix B 346...
Страница 366: ...DMTA 10040 01EN Rev E February 2018 List of Figures 356...
Страница 368: ...DMTA 10040 01EN Rev E February 2018 List of Tables 358...