DMTA-10040-01EN, Rev. E, February 2018
Chapter 4
82
4.3
Menus
The NORTEC 600 menus described in this section are opened by pressing the
corresponding menu keys, which are described in “Menu Keys” on page 77.
4.3.1
Frequency (FREQ 1) Menu — MAIN FILTER Key
FREQ (frequency)
The
FREQ
setting determines the frequency of the drive signal of the eddy
current probe. This is adjustable from 10 Hz (0.01 kHz) to 12 MHz.
To adjust the frequency setting, press the MAIN FILTER menu key (
),
followed by the A key. With
FREQ
highlighted, rotate the knob until the desired
frequency is displayed.
To speed up the frequency selection process, press the Enter key (
) once while the
FREQ
function is highlighted to enable the coarse knob function.
FREQ
will be
underlined when this function is enabled. To turn off the coarse-knob function, press
one more time.
ANGLE (rotation)
The phase angle (or rotation) of the eddy current signal is set using the ANGLE
key (
). By default, the angle is set in 1 degree increments, from 0 to
359 degrees.
To adjust the angle setting, press the MAIN FILTER menu key (
), followed by
the B key. With
ANGLE
highlighted, rotate the knob until the desired angle is
displayed.
TIP
Содержание nortec 600
Страница 8: ...DMTA 10040 01EN Rev E February 2018 Table of Contents viii...
Страница 16: ...DMTA 10040 01EN Rev E February 2018 Labels and Symbols 6...
Страница 30: ...DMTA 10040 01EN Rev E February 2018 Introduction 20...
Страница 58: ...DMTA 10040 01EN Rev E February 2018 Chapter 1 48...
Страница 71: ...DMTA 10040 01EN Rev E February 2018 Software User Interface 61 Press the Return key to exit...
Страница 72: ...DMTA 10040 01EN Rev E February 2018 Chapter 2 62...
Страница 342: ...DMTA 10040 01EN Rev E February 2018 Chapter 7 332...
Страница 356: ...DMTA 10040 01EN Rev E February 2018 Appendix B 346...
Страница 366: ...DMTA 10040 01EN Rev E February 2018 List of Figures 356...
Страница 368: ...DMTA 10040 01EN Rev E February 2018 List of Tables 358...