DMTA-10040-01EN, Rev. E, February 2018
Using the Instrument
245
Figure 5
‑
174 The tube scan result
13. When you are satisfied with the setup, press the REF SAVE (
) key.
The file will be saved with a time/date stamp in the memory menu. This stored file
can be recalled and used later to show additional features.
To verify the setup with the support ring over a flaw
For this subprocedure, the support ring is moved directly over a flaw in order to
verify that the flaw is still visible with the
MIX
display.
1.
Slide the o-rings on the tube to allow movement of the support ring. Move the
support ring over the flaw and place the o-rings against the support ring to keep
it in position (see Figure 5-175 on page 246).
NOTE