DMTA-10040-01EN, Rev. E, February 2018
Using the Instrument
235
Figure 5
‑
159 Example of lower signal lobe when pulling the probe over a flaw
Figure 5
‑
160 The scan signal of the thru
‑
wall hole
3.
Press the ANGLE key (
), and then rotate the frequency 1 signal (green) until
the hole signal reaches a phase of about 40° (see Figure 5-161 on page 236).
Choose the polar (
WEB
) grid shown in this example to make it easier to adjust the
phase angle and gain. (For details on grid selection, see “
major graticule represents 20° and 10 % of full-screen height.
TIP