Olympus MV-1 Скачать руководство пользователя страница 18

MV-1/MV-2 Endoscope Disinfectors

Frequently

Asked

Questions

09FQ0015 18/18

Two Corporate Center Drive, Melville, NY 11747

Copyright 2001, Olympus America Inc.

Copyright 2001 Olympus America Inc. This document and the contents contained herewith remain the property of Olympus America Inc. Reproduction of this document or any part

thereof by any means (electronic or otherwise) without the expressed written consent of Olympus America Inc. is strictly prohibited. Only printed copies with "Controlled Document -- Do

Not Copy" stamped in red are considered controlled copies. All other printed copies are considered Uncontrolled and may not contain the latest document revisions.

Troubleshooting: MV-1/MV-2 Control Panel (Cont’d)

#

 

Ensure the MV-1/MV-2 control panel is operating properly. If necessary, replace the MV-1/MV-2 control
panel.

#

 

Check the ribbon cables within the MV-1/MV-2 to be sure that they are connected properly and are not
loose.

Troubleshooting: Returning the Unit for Repair

If the MV-1/MV-2 Endoscope Disinfectors must be returned to Medivators, Inc. for repair or replacement,
contact Olympus by calling telephone # 1-800-848-9024. Choose the appropriate prompt for connection to
Medivators, Inc.

A “Hot Swap” will be performed, which means that Medivators, Inc. will provide the facility with a loaner unit to
replace the disabled MV-1/MV-2 until the unit is either repaired or replaced. Medivators, Inc. will provide
shipping instructions.

Содержание MV-1

Страница 1: ...tronic or otherwise without the expressed written consent of Olympus America Inc is strictly prohibited Only printed copies with Controlled Document Do Not Copy stamped in red are considered controlled copies All other printed copies are considered Uncontrolled and may not contain the latest document revisions Troubleshooting Rinse Alarm Model Overview Cabling and Connectivity Compatible Component...

Страница 2: ...r rinses follow the high level disinfection cycle to provide maximum rinsing The high level disinfectant reservoir allows visual inspection of the disinfectant easy access for disinfectant testing and is removable for cleaning The disinfectors are built without open vents and operate as closed systems to allow minimal fume emission Data reports that less than 0 04ppm glutaraldehyde vapor is emitte...

Страница 3: ... other printed copies are considered Uncontrolled and may not contain the latest document revisions Model Overview Cont d Additional Components The following components are supplied with the MV 1 MV 2 Endoscope Disinfectors MV 1 MV 2 User service manual Accessory bag 1 Alcohol flushing adapter Water hose Drain hose Power plug Endoscope connector standard set for Olympus or Pentax or Fujinon flexib...

Страница 4: ...of Olympus America Inc is strictly prohibited Only printed copies with Controlled Document Do Not Copy stamped in red are considered controlled copies All other printed copies are considered Uncontrolled and may not contain the latest document revisions Compatible Components The MV 1 MV 2 Endoscope Disinfectors are compatible with the following components Mobile Cart CLM 1005 LCG transfer pump CLM...

Страница 5: ...1 Connection of the Drain to the MV 1 MV 2 Flexible drain hose Connect the flexible drain hose to the white drain fitting on the back of the MV 1 MV 2 and attach the screw clamp Caution The drain must be 12 in below the drain fitting on the disinfector and loops in the drain hose must be eliminated to ensure that fluids drain thoroughly Incomplete draining will result in dilution of the LCG and or...

Страница 6: ...lace Place the white cover over the drain screen It should fit behind the drain bar Caution Do not attempt to lift up the MV 1 MV 2 using the drain bar Damage to the MV 1 MV 2 may result 5 Connection of the Channel Connectors to the MV 1 MV 2 Refer to the MV 1 MV 2 Connector Matrix for a listing of endoscopes and their corresponding channel connectors Channel Connectors Attach the channel connecto...

Страница 7: ...e hose fitting water inlet on the back of the MV 1 MV 2 9 Installing the Filters in the Water Filtration System CLM 1076 Filters Prefilter CF01 0017 Postfilter CF01 0018 or Prefilter and Postfilter Set CF01 0007 Caution Personal protective equipment should be worn to guard against reprocessing chemicals and potentially infectious material when performing this procedure Wear appropriate equipment s...

Страница 8: ...of the filter and are centered when the filter is inserted into the housing If the postfilter has been removed place the new postfilter into the blue postfilter housing with the black O ring facing UP and the closed end facing DOWN Make sure the O ring is securely fastened to the peg at the top of the housing It may be necessary to moisten the O ring with water or a water based lubricant 10 Connec...

Страница 9: ...rs User Service Manual All endoscopes must be precleaned before placement in the MV 1 MV 2 Follow the manufacturer s instructions and established professional guidelines see page 5 in the MV 1 MV 2 Endoscope Disinfectors User Service Manual for further instructions on the cleaning steps that must be performed prior to placing an endoscope into the MV 1 MV 2 Caution Effective reprocessing can only ...

Страница 10: ...parts and filters must be used to assure proper operation and efficiency of the MV 1 MV 2 Do not use gauze cotton swabs or other lint producing cleaning aids on the MV 1 MV 2 These materials shed fibers that can become caught in the channels of endoscopes or in the components of the MV 1 MV 2 Do not use products that contain alcohol when cleaning the lid of the MV 1 MV 2 Alcohol may cause crazing ...

Страница 11: ... 1 MV 2 Water Flow A channel connector should be connected to each port in the MV 1 MV 2 basin Turn the MV 1 MV 2 power switch ON The power switch is located on the right side of the MV 1 MV 2 Press START on the MV 1 MV 2 control panel Wait for the pump to feed water through the channel connectors Confirm that water is flowing properly through each connector Note Flow volume will vary depending up...

Страница 12: ...nimum of 4 gallons 15 1 L of LCG should be in the LCG reservoir of the MV 1 A minimum of 5 gallons 18 9 L of LCG should be in the LCG reservoir of the MV 2 Note Recommendations for the disposal of LCG vary between formulations and locations Follow the policies and procedures applicable to the facility Before Operating LCG Concentration The LCG must be tested daily to ensure that it contains an ade...

Страница 13: ...he system to operate properly The MV 1 MV 2 disinfectors require a minimum incoming water pressure of 40 psi 2 8 bar as indicated by the reading on gauge 1 Check the gauges located on the water filtration system When the reading on gauge 1 is normal and the reading on gauge 3 drops below 40 psi it is likely that one or both of the filters must be changed Read the gauges to determine which filter m...

Страница 14: ...n occur and possible solutions to these problems Some of the solutions given may require the intervention of a trained service engineer If this is the case contact Olympus The following documents are referenced in this section Click on the document title below to open the document MV 1 MV 2 Endoscope Disinfectors User Service Manual Document 09 WI 0240 MV 1 MV 2 Endoscope Disinfectors Confidence T...

Страница 15: ...bleshooting LCG 1 Was the LCG lost through the overflow drain Check to see if the MV 1 MV 2 is level If necessary adjust the legs of the MV 1 MV 2 Make sure any material or debris is not covering the float tower Make sure that the float sensor is not defective Replace the float sensor if necessary 2 Is the LCG not completely filling the basin Make sure the LCG tubing does not contain any kinks or ...

Страница 16: ...h If drainage is too slow contact Olympus Check the Solution Out solenoid to be sure it is not leaking water This applies only to the LCG reservoir overfilling problem If there is water leakage contact Olympus 5 Is there loss of LCG during MV 1 MV 2 operation If LCG is lost via the drain contact Olympus Check the drain screen to ensure that the drain is not blocked Make sure the LCG OUT quick conn...

Страница 17: ...fully inserted Ensure the check valve that is in line with the alcohol manifold is not leaking air If this is the case replace the check valve Make sure the air pump is not defective If necessary replace the air pump Check the internal air pump tubing to be sure it is not loose If necessary reconnect the tubing properly If air is flowing instead of fluid the motherboard might be defective If neces...

Страница 18: ...ed and may not contain the latest document revisions Troubleshooting MV 1 MV 2 Control Panel Cont d Ensure the MV 1 MV 2 control panel is operating properly If necessary replace the MV 1 MV 2 control panel Check the ribbon cables within the MV 1 MV 2 to be sure that they are connected properly and are not loose Troubleshooting Returning the Unit for Repair If the MV 1 MV 2 Endoscope Disinfectors m...