![Olympus Magna-Mike 8600 Скачать руководство пользователя страница 89](http://html1.mh-extra.com/html/olympus/magna-mike-8600/magna-mike-8600_user-manual_3943437089.webp)
DMTA-10026-01EN, Rev. D, January 2015
Basic and Multipoint Calibration
79
Place the target ball into the BALL ON FIXTURE before placing the fixture onto the
probe. Dropping the ball repeatedly into the fixture as it rests on the probe may pit the
end of the wear tip, causing small measurement inaccuracies.
Figure 4-6 Alignment of target ball in fixture on probe
When using the target disk and the chisel wear tip, make sure that the disk is aligned
perpendicularly to the chisel edge (see Figure 4-7 on page 80).
NOTE
Target ball
Probe
BALL ON FIXTURE
IMPORTANT
Содержание Magna-Mike 8600
Страница 7: ...DMTA 10026 01EN Rev D January 2015 Table of Contents vii List of Tables 169 Index 171...
Страница 8: ...DMTA 10026 01EN Rev D January 2015 Table of Contents viii...
Страница 10: ...DMTA 10026 01EN Rev D January 2015 List of Abbreviations x...
Страница 28: ...DMTA 10026 01EN Rev D January 2015 Introduction 18...
Страница 70: ...DMTA 10026 01EN Rev D January 2015 Chapter 3 60...
Страница 112: ...DMTA 10026 01EN Rev D January 2015 Chapter 5 102...
Страница 120: ...DMTA 10026 01EN Rev D January 2015 Chapter 6 110...
Страница 146: ...DMTA 10026 01EN Rev D January 2015 Chapter 7 136...
Страница 160: ...DMTA 10026 01EN Rev D January 2015 Chapter 8 150...
Страница 164: ...DMTA 10026 01EN Rev D January 2015 Chapter 9 154...
Страница 168: ...DMTA 10026 01EN Rev D January 2015 Appendix A 158...
Страница 178: ...DMTA 10026 01EN Rev D January 2015 List of Figures 168...
Страница 180: ...DMTA 10026 01EN Rev D January 2015 List of Tables 170 Table 31 Calibration kits 163...
Страница 186: ...DMTA 10026 01EN Rev D January 2015 Index 176 WINXL about 137...