2.5 Operating and storage conditions

Always meet the operating and storage conditions.

Operating environment

Ambient temperature

+5 °C - +40°C

Relative humidity

5% - 85% (without condensation)

Altitude

Up to 2000 m

Storage conditions

Ambient temperature

-20°C - +60°C

Relative humidity

10% - 90%

Camera storage

When not using the camera, attach the cover to the C-mount and
store the camera in a safe place.

2 The OLYMPUS LC35 camera

11

Содержание LC35

Страница 1: ...Instruction manual OLYMPUS LC35 Digital Microscope Camera Optical Microscope Accessory English...

Страница 2: ...onditions 11 3 System environment 12 3 1 Microscope and camera adapter 12 3 2 Recommended computer and monitor configuration 13 3 3 Restrictions on use 13 4 Assembly 15 4 1 Attaching the camera 15 4 2...

Страница 3: ...al This instruction manual is for the OLYMPUS LC35 microscope digital camera Information about the operation of the microscope computer or software can be found in the instruction manuals for the corr...

Страница 4: ...tended to be used for the acquisition of digital images but not for medical diagnostic purposes 1 2 Avoiding personal injury and device damage WARNING Avoiding personal injury Electric shock due to da...

Страница 5: ...nly the supplied USB camera cable Operate the camera only with the supplied USB camera cable Only then will the functionality of the camera and compliance with the EMC basic standards be ensured Don t...

Страница 6: ...ions see page 11 1 3 Explanation of the symbols used WARNING Indicates a potentially hazardous situation which if not avoided may result in major injury CAUTION Indicates a potentially hazardous situa...

Страница 7: ...OLYMPUS LC35 camera 2 1 Scope of supply Camera USB camera cable Safety and warning information A printed copy of the safety and warning information are supplied with the camera 2 The OLYMPUS LC35 cam...

Страница 8: ...er manuals 2 2 System diagram To use the OLYMPUS LC35 with a light microscope you need suitable software and a computer as well as a monitor For the recommended computer and monitor configuration refe...

Страница 9: ...us ims com service and support downloads 2 3 LED status indication There is an LED on the camera which indicates the following statuses LED Status The LED flashes green The camera is being initialized...

Страница 10: ...ng area 5 702 mm x 4 277 mm 7 13 mm diagonal Maximum resolution 2160 x 1620 pixels 3 5 million pixels Camera port of the microscope C Mount Input output On the camera USB 3 1 type C On the computer US...

Страница 11: ...Ambient temperature 5 C 40 C Relative humidity 5 85 without condensation Altitude Up to 2000 m Storage conditions Ambient temperature 20 C 60 C Relative humidity 10 90 Camera storage When not using t...

Страница 12: ...dapter connected to OLYMPUS LC35 U TV1XC U TV1X U CMAD3 O U TV0 63XC O U TV0 5XC 3 O U TV0 35XC 2 0 5XC integrated in SZ61TR O O Compatible Limited compatible Depending on the microscope used this cam...

Страница 13: ...rt USB 3 1 type A Removable media drive Optical drive double layer capable Computer input device Recommendation 3 button mouse with wheel Keyboard Operating system Microsoft Windows 10 64 Bit 3 3 Rest...

Страница 14: ...o long see following example clouding can occur at the edges of the field of view Examples of an intermediate tube that is too long n Two or more intermediate adapters are being used together n BX3 se...

Страница 15: ...amera adapter to the camera port of the microscope Checking the camera s orientation If the camera is not attached in the correct orientation the camera image does not match the image visible through...

Страница 16: ...a cable Only then will the functionality of the camera and compliance with the EMC basic standards be ensured 1 Connect the USB type C plug a to the camera s USB type C socket b 2 Connect the USB came...

Страница 17: ...OLYMPUS Stream PRECiV Read the installation manual before installing the software TWAIN driver DP2 TWAIN LCmicro This software is available only as a download from the Olympus Internet site https www...

Страница 18: ...5 Acquiring images You can find detailed instructions in the instruction manual or in the Help of the software you are using Example of the acquisition process OLYMPUS LC35 18...

Страница 19: ...il smudges on the protective glass carefully clean it with a clean lens cloth moistened with pure alcohol Cleaning the housing Parts other than the glass components should be cleaned with a clean clot...

Страница 20: ...a USB 2 0 port Connect the camera to an USB 3 1 port Problem Live image is not displayed Possible cause Remedy The illumination of the microscope is not turned on Switch on the microscope illumination...

Страница 21: ...save the images Close the loaded images Then acquire the next snapshot Problem The image is too bright Possible cause Remedy The exposure compensation is set too high to the side Set the exposure com...

Страница 22: ...enuation filters The manual exposure time is set too low Increase the manual exposure time Problem The colors of the image are inaccurate Possible cause Remedy The reference area for white balance was...

Страница 23: ...aperture and focus the microscope on the sample The field diaphragm is too wide open Narrow the field aperture appropriately and focus the microscope on the sample Lens components of the microscope a...

Страница 24: ...your local Olympus support if you have any questions about the product More information can be found on this website https www olympus lifescience com support service https www olympus ims com de serv...

Страница 25: ...waste but should be collected separately Refer to your local authority in the EU for return and or collection systems available in your country RoHS conformity Europe This Olympus Soft Imaging Solutio...

Страница 26: ...ular installation If this equipment does cause harmful interference to radio or television reception which can be determined by turning the equipment off and on the user is encouraged to try to correc...

Страница 27: ...ce may not cause harmful interference and 2 this device must accept any interference received including interference that may cause undesired operation Responsible Party Name Olympus Scientific Soluti...

Страница 28: ...China RoHS conformity China Pb Hg Cd Cr VI PBB PBDE X O O O O O X O O O O O X O O O O O O O O O O O SJ T 11364 GB T26572 X GB T26572 OLYMPUS LC35 28...

Страница 29: ...t Korea Co Ltd Seocho dong Seocho Central IPARK 5th 6th Floor 36 Banpo daero 18 gil Seocho gu Seoul Evident Scientific Singapore Pte Ltd 438B Alexandra Road 03 07 12 Alexandra Technopark Block B Lobby...

Отзывы: