SUCTION PUMP 

KV-6

 

Specification

Classification
(electromedical 
equipment)

Standards compliance

Complies with EN ISO 10079-1 and EN IEC 60601-1 and UL 60601-1

Electro-magnetic 
Compatibility

This product complies with the requirements of EN IEC 60601-1-2 for emissions and immunity, and as such, its operation is unlikely to 
be affected by, or cause interference with, equipment meeting appropriate EMC standards. As a precaution, equipment which may be 
sensitive to interference outside the limits specified by EN IEC 60601-1-2 should not be placed in close proximity to the KV-6.

Type of protection against
electrical shock

Classification according to EN IEC 60601-1 / UL 60601-1: Class 1 with Type CF applied part.

Degree of protection
against electrical shock

In accordance with EN IEC 60601-1 and UL 60601-1, the KV-6 is marked with the symbol to indicate the provision of an adequate 
degree of protection against electrical shock and that it has an applied part isolated from all other parts of the equipment.

Degree of protection
against explosion

None: the Olympus KV-6 must NOT be used in the zone of risk of flammable anaesthetic gases.

Mode of operation

Continuous

Regulatory status

European Economic Area
(EEA)

CE This mark on the product indicates compliance with Directive 93/42/EEC relating to medical devices, Class IIa.

Dimensions and weight

Dimensions

Height: 210mm

Width: 305mm

Depth: 375mm

Weight

9.3 kg

Power switch

Marking

The power switch is marked  . A white LED in the center signalizes it is turned on

Power requirements

Power supply

100-240 V~

Fusing

2 x T1.6 AH

Marking:

 The mark ~ on  

the product indicates the 
requirement for an AC  
power supply

Frequency

50/60Hz

Power rating

120 W

Environmental 
conditions

Ambient Temperature

Operational:

+5ºC to +40ºC (+41ºF to +104ºF)

Storage:

-20ºC to +50ºC (-4ºF to +122ºF)

Relative humidity

Operational:

30-75 % 

Storage:

20-95%

Atmospheric pressure

Operational/Storage:

70.0 to 106 kPa

Fluid Ingress

In accordance with EN IEC 60601-1/UL 60601-1, the product is marked IP21 to indicate that it is provided with an enclosure that prevents entry of such an amount of 
falling liquid as might interfere with its safe and satisfactory operation when correctly positioned.

Resistance to 
chemicals

The external surfaces of the Olympus KV-6 are resistant to: 2% aqueous neutral detergent, 70% ethyl alcohol, isopropyl alcohol, water.

Pump specification

Nominal vacuum:

 95 kPa

Nominal free air flow rate:

 50 l/min, adjustable to 40 or 60 l/min.

In accordance with ISO 10079-1, the unit is marked with the words ‘HIGH VACUUM, HIGH FLOW’ to indicate it has a vacuum of at least 60 kPa in 10 seconds and a 
free air flow rate of 20l/min or greater.

Thermal cut-out

The temperature is observed by a sensor in the housing. If the temperature in the housing exceeds 80 °C (176 °F), the software stops the motor and turns on the yellow 
LED. It is possible to restart the motor again as soon as the temperature in the housing drops below 70 °C (158 °F).

Single use suction liner

To prevent fluid from being drawn up to the pump the single use suction liner contains a mechanism to shut off the flow when the liner is overfilled. This mechanism is 
activated when contact is made with aspirated fluid.

Suction jar

The suction jars can be autoclaved up to 135 ºC or disinfected. The jar is impact resistant.

Suction jar lid

The suction jar lid clips and where applicable the float mechanism, are autoclavable up to 135 ºC. The inlet connection (from the endoscope) is marked with the word 
PATIENT, the outlet connection (to the pump) is marked with the word VACUUM.

Tubing

Use only tubing specified in the KV-6 Instructions for Use, otherwise equipment malfunction may result.

Vacuum gauge

Range:

 0 to 100 kPa

Accuracy:

 ±2.5%

Printed in UK PL022/201505