DMTA034-01EN, Rev. B, March 2014
Scanner Assembly and Setup
33
9.
Install the new probe onto the wedge.
Before installing a new probe on the wedge, ensure that there is a sufficient amount of
couplant between the probe and the wedge.
10. Tighten the screws that hold the probe on the wedge.
Do not overtighten the screws: overtightening might crack the wedge.
11. Reinstall the yoke on the wedge.
12. Reinstall the probe assembly on the spring-loaded arm.
This concludes the procedure for replacing a phased array probe.
2.4
Changing the Frame Bar Sections
When necessary, the frame bar sections of the scanner can be replaced with sections of
a different length to suit physical constraints or probe-configuration requirements.