Chapter 5 Troubleshooting
75
EVIS EXERA GIF/CF/PCF TYPE 160 Series OPERATION MANUAL
When the endoscopic image does not appear on the monitor
or the frozen image cannot be restored
1.
Turn OFF all equipment except the video system center, the light source
and the monitor.
2.
Turn the video system center and light source OFF and then ON again. If
the endoscopic image appears or the frozen image is restored, follow the
procedure of Step 2. and below “When the endoscopic image appears on
the monitor” on page 74.
When the endoscopic image still does not appear or the frozen image
cannot be restored, perform the following steps.
3.
Turn OFF the video system center, the light source and the monitor.
4.
When using an endo-therapy accessory, withdraw the endo-therapy
accessory slowly while the tip of the endo-therapy accessory is closed
and/or retracted into its sheath.
5.
When using an endoscope with the flexibility adjustment function, set the
insertion tube to the softest condition (for CF-Q160AL/I, PCF-160AL/I only).
6.
Turn the UP/DOWN and RIGHT/LEFT angulation locks to the “F
”
direction to release them.
7.
Turn the UP/DOWN and RIGHT/LEFT angulation control knobs to their
respective neutral positions (see Figure 3.6). Release the angulation control
knobs and carefully withdraw the endoscope.
8.
When the splinting tube is used, withdraw it from the patient’s anus (for CF
models only).
9.
Remove the mouthpiece from the patient’s mouth (for GIF models only).