4.10 Switching OFF Protocol 

Check through TLL microscopy resource booking if there is a user immediately after you. If 
there is a user immediately after you, leave the system on and log off the PC. Otherwise 
switch off the system. 

(a)

 

Save your images through TLL common drive folder. Images saved in computer D/E 
drives will be 

deleted within a month without prior warning

.  

(b)

 

Lower down the objective lens completely. 

(Z value= 0) 

(c)

 

Remove the specimen holder with the sample and return the stage to central 
position. 

(d)

 

Carefully clean the objective lens with Whatman lens tissue using 100% ethanol. 

(e)

 

Switch lens back to 10x/0.3 objective lens. 

(f)

 

Exit the FV31S-SW software and shut down PC 

(2)

(g)

 

Press “OFF” on the touch screen panel and wait for 5 secs for the light to disappear. 

Then, switch OFF the power button at the back of touch screen panel 

(6)

(h)

 

Switch off the Stage controller box 

(3)

, CBH controller 

(4)

(i)

 

Switch off the HBO/Mercury lamp if used 

(5)

. (Record down the lifespan hours in the 

logbook.) 

(j)

 

Switching OFF lasers,    

 

To turn off 405/488/561/640nm, turn off the individual laser lines and turn the 

key to “standby”, 

(8)

. Switch off the main combiner 

(7)

.   

 

To turn off 445/514nm, turn off the individual laser lines and turn the key to 

“standby” 

(10)

. Switch off the sub combiner 

(9)

.   

(k)

 

Turn off the incubator 

(12)

 if it was on. 

(l)

 

Finally, Switch off the master switch on the wall 

(1)

(m)

 

Fill in the logbook with the objective lens used. 

 

 

 

 

 

 

Содержание FV3000

Страница 1: ...oking and System Access 4 3 1 Acknowledgements 5 4 Operation Procedures 5 4 1 Switching ON Protocol 5 4 2 Objective lens 6 4 2 Software Initiation 7 4 3 Setting up directory to save images 8 4 4 Laser...

Страница 2: ...Immersion Oil Modern immersion oil has no known hazards to human beings so far yet they can cause discomfort to a person if the immersion oil has been left on skin for too long or inhaled Therefore pl...

Страница 3: ...taff before they are allowed to access the confocal system 2 1 To arrange a training session Please email Bioimaging Facility bioimaging tll org sg for a training session Fill in the necessary particu...

Страница 4: ...er 6pm Sat Sun Public holidays If users require more slots in a particular week they can book on weekdays during non peak hours after 6pm on weekends Sat and Sun and on public holidays If extra time i...

Страница 5: ...RT TIMING b Switch on the Master switch on the wall 1 PC 2 Wait until windows icon pops up c Switch on Stage controller box 3 CBH controller 4 d Optional Turn on the Metal Halide lamp 5 if needed e Tu...

Страница 6: ...t to 10x objective or lower then click YES Cleaning the stage will take some time b Otherwise click NO to skip cleaning 4 2 Objective lens Objective lens Immersion media 1 25x 0 04 Plan Apochromat Air...

Страница 7: ...n the stage mechanism This adds time to the initialization process Do not place anything on the stage mechanism until the initialization process is complete If you are not the first user of the week C...

Страница 8: ...y to save all your images Please save in your folder You have several different options on where you can save your images Option 1 D or E Users shaalini create your own folder BPAE cells oir Do keep i...

Страница 9: ...oduce intense background noise SD keep HV within 600 800 More than 800 will contribute to intense background noise Under PMT settings tab Mode Select VBF Average average a given number of scans to get...

Страница 10: ...to zoom images only in the center Ideal zoom which satisfy the Nyquist limit can be achieved by zooming till the red arrow any zoom beyond it will result in poor resolution and result in more photoble...

Страница 11: ...the laser power until red pixels disappear Increase Offset until a dusting of blue pixels in the background are observed Click LSM start to capture an image 4 7 How to do a Z stack for a 3D Image xyz...

Страница 12: ...your timelapse experiment Interval FreeRun continues to capture images until it has completed your selected cycle You could always choose an interval of your choice 4 9 Reuse previous settings and ex...

Страница 13: ...h the sample and return the stage to central position d Carefully clean the objective lens with Whatman lens tissue using 100 ethanol e Switch lens back to 10x 0 3 objective lens f Exit the FV31S SW s...

Страница 14: ...ission please contact Bioimaging Group No stage movement Check the coarse fine movement setting near the focus and xy position knobs It is a physical button that has no electronic indication elsewhere...

Страница 15: ...id white or coloured light sometimes red light might be indicative of error III Restart the main controller right at the bottom if none of the steps taken work especially if the machine has been left...

Страница 16: ...and cannot be recovered Restart computer terminal Startup popup error code LAST_OPERATION_ NONCOMPLETED The last user did not shut down the program properly or the program encountered a fatal error du...