TROUBLESHOOTING GUIDE
TROUBLESHOOTING
1 - 2
Page
Phenomenon
Cause
Treatment
Manual Ref. Pages
9. FLUOVIEW software
cannot be booted up.
Something wrong with the
‘ini’ file.
1.Terminate FLUOVIEW
software and FLUOVIEW
Setup.
2.Delete [GBSTATE.ini] file
in [FLUOVIEW System]
folder in HDD where
FLUOVIEW soft ware is
installed, Then set it up
again by FLUOVIEW Setup
program.
3.Boot up FLUOVIEW
software.
“Maintenance
Section 1-3”