WARNING: 

This guide is only a summary of the steps necessary to properly disinfect your endoscope. Be sure to follow the 

detailed steps outlined in the 

ENDOSCOPE REPROCESSING MANUAL 

that was included with your endoscope when purchased. 

To use this chart as a wallchart, cut  
off the top of the chart as illustrated.

To insert this chart into a binder, punch out 
the holes and fold the chart as illustrated.

Suction Cleaning Adapter

Tube to suction

To Suction Pump

Suction Cleaning

Adapter (MAJ-222)

To Suction Pump

Clean the bronchoscope 

meticulously prior to disinfection.

Single-Use Suction Valve (MAJ-209)

Suction Cylinder

Single-Use Biopsy Valve (MAJ-210)

Remote Switches

Universal Cord

Instrument/Suction Channel Port

Electrical Connector

Water-Resistant Cap 

(MH-553)

Venting Connector

Distal Tip

Bending Section

Light Guide

Boot

UP/DOWN Angulation Control Lever

Control Section

Insertion Tube

Alcohol

HIGH-LEVEL DISINFECTION

RINSING

TRANSPORTATION

DRYING

NOMENCLATURE

IMPORTANT

1

2

3

4

5

6

9

10

7

8

©

 2

0

0

4

 O

ly

m

pu

A

m

er

ic

In

c.

 /

 P

ri

nt

ed

 in

 U

SA

  

0

1

/0

4

 –

 P

W

C

-1

0

0

0

1

6

EXERA

 BF-160 Series and EVIS

 BF-240 Series Bronchovideoscopes

BRONCH

OSC

O

PE DISINFECTION GUIDE

 Refer to the companion 

BRONCHOSCOPE CLEANING GUIDE

 for detailed cleaning information

WARNING:

 

This guide is only a summary of the steps necessary to properly disinfect your bronchoscope. Be sure to follow the 

detailed steps outlined in the 

ENDOSCOPE REPROCESSING MANUAL 

that was included with your bronchoscope when purchased. 

Attach a syringe to the adapter. 

Pull the plunger of the syringe to

fill the instrument/suction channels

and channel cleaning adapter

with disinfectant solution.

Fill basin with disinfectant.  Attach a 

suction cleaning adapter (MAJ-222) 

to the previously cleaned broncho-

scope.  Immerse completely in 

disinfectant solution.

While immersed, detach the adapter 

from the bronchoscope. Use a clean, 

lint-free cloth to wipe away any bub-

bles clinging to external surfaces.

Soak all equipment according to

the time and temperature 

recommended by the disinfectant 

manufacturer. Use a timer to 

measure the exact contact time.

Connect the suction cleaning

adapter and syringe. Inject air 

to remove disinfectant from the 

instrument/suction channels, 

then remove the bronchoscope 

from the disinfectant. 

Disconnect the suction cleaning

adapter and immerse the 

bronchoscope and adapter in 

sterile water. Using a sterile, 

lint-free cloth, thoroughly rinse 

and wipe all surfaces.

Connect suction cleaning adapter 

to bronchoscope and to a tube 

from a suction pump. Aspirate 

water through channels for 30 

seconds. Remove from water and 

aspirate air through channels for 

60 seconds.

Turn the suction pump off. Hold the control 

section with the openings pointing down and 

disconnect the suction cleaning adapter.

Dry the bronchoscope and adapter with an 

alcohol-moistened cloth. Make sure that the 

equipment is completely dry prior to storage.

Hold all parts of the endoscope when carrying. Do 

not let the ends of the bronchoscope swing freely.

• Wear appropriate personal protective 

equipment during reprocessing.

• Autoclave all autoclavable accessories.
• Store the bronchoscope hanging vertically 

with all valves and caps removed.

• Use a compatible disinfectant and test

it for proper concentration as recom-

mended by the manufacturer.

• Do not use the carrying case to store 

"patient-ready" endoscopes. Any broncho-

scope placed in the carrying case must be 

reprocessed prior to patient use.

Immerse the bronchoscope's distal 

end in 70% alcohol and aspirate 

alcohol for 5 seconds. Remove the 

distal end from the alcohol and 

aspirate air for 20 seconds.