Olympus EU-ME2 Скачать руководство пользователя страница 1

VIDEO SOURCE key

Image on monitor

LCD touch panel

QUICK-REFERENCE GUIDE – BASIC CONTROLS

Insert mechanical scanning scopes in the left socket (SOCKET1) 
or electronic scanning scopes in the right socket (SOCKET2) 
and turn the lever clockwise to tighten the endoscope connector.

Press the power switch and confirm that the power indicator 
(green light) is lit.

Press the “

B

,” “

FLOW,

” and “

PW

” keys on the keyboard to select B 

mode, FLOW mode, and Pulse Wave Doppler Mode.

Press the “

CH-EUS

” and “

ELST

” buttons to select between

CH-EUS mode and Elastography Mode.

Press the 

ID INPUT

 key and the 

Patient Data

 screen appears 

on monitor (above).

Enter all the necessary data and click 

OK

 to proceed.

Connecting Ultrasound Endoscopes

Mode Selection

Video Source

ID Input

1

1

1

2

2

2

*  Please confirm either the power switch or

 ACTIVE

 switch is turned off 

when the scope is being disconnected.

* It takes approximately one minute to start up EU-ME2.
*  You can change scopes by turning off the 

ACTIVE

 switch, which has 

eliminated the hassle of turning off EU-ME2.

*  Switching is not possible in FREEZE Mode. Select modes in live mode.
*  Refer to the separate sheet regarding FLOW mode selection (COLOR 

FLOW, POWER FLOW, and H-FLOW).

*  The Mode menu on the LCD touch panel is displayed at the upper right 

corner of each main screen (shown after unfreezing).

*  The Mode menu on the LCD touch panel may not appear in some models.

* ID input is mandatory to save data after examinations.
*  ID input is not available when using the CV-290, CV-190, or CV-180. 

ID input should be made through these-video system centers.

Press the 

VIDEO SOURCE

 key on the keyboard and switch US and EVIS.

* EVIS operation can be controlled by the LCD touch panel on the keyboard.

With mechanical scanning scopes

ID INPUT key

Patient Data screen on the monitor

With electronic scanning scopes

Power switch

ACTIVE switch

US mode

EVIS mode

Switch

MODE buttons (LCD)

MODE keys

Contrast Harmonic EUS

Pulse Wave Doppler

FLOW mode

Elastography

B mode

Socket1

Lever

Ultrasound connector

Lever

Ultrasound connector

Socket2

Отзывы: