background image

ESG-150 Electrosurgical Generator

CAUTION: 

 This guide is only a summary of the instructions for use of the ESG-150 Electrosurgery Generator. Be sure to reference the instruction 

manual that was included with your product purchase.

ESG-150 User Flexibility

Bipolar

Neutral

On/Off

Monopolar

Graphical 

User 

Interface

CONTINUED ON NEXT PAGE

ESG-150

 (

Electrosurgical Generator)

 

Quick Reference Guide

Basic Level

Advanced Level

Содержание ESG-150

Страница 1: ...G 150 Electrosurgery Generator Be sure to reference the instruction manual that was included with your product purchase ESG 150 User Flexibility Bipolar Neutral On Off Monopolar Graphical User Interfa...

Страница 2: ...ctive Socket Active Socket Active Socket Contact Quality Monitor Contact Quality Monitor Change Output Socket Change Output Socket Change Output Socket Change Output Socket System Setup System Setup S...

Страница 3: ...Guide Select Basic or Advanced Level Functionality 1 Select the Menu button 2 Select User Level 3 Select Basic or Advanced Functionality 4 Select the OK button to confirm 5 Functionality has changed o...

Страница 4: ...fined Settings Mode menu selection 1 Select the mode by tapping on the desired mode located on the bottom of the touch screen 1 Select the Menu 1 Select the mode by tapping on color bar on the touch s...

Страница 5: ...ce Guide ESG 150 System Operating Steps Changing the Power Level Changing the Effect 4 5 Select to adjust power level Select to adjust effect Effect is only available in the Advanced Level Function 6...

Страница 6: ...trode lights up and alarm tone sounds Is the neutral electrode correctly connected to the device Does the neutral electrode have sufficient contact to the skin Correct Green Error Red Monitor Neutral...

Страница 7: ...er Defined Settings STEP 1 On the home screen Select Preferred Modes Select Power Level Select Effect Level Select Menu STEP 2 Settings Setup Screen Select Save STEP 3 Save Setting Screen Enter the se...

Страница 8: ...based upon previous general clinical usage of the device Please note that clinical staff are solely responsible for selection of an appropriate mode and power level dependent upon the instrument being...

Страница 9: ...PulseCut ForcedCoag SoftCoag Continuous cutting Controlled cutting with an intermittent duty cycle of different duration Superficial pinpoint coagulation Deeper coagulation High Power Cut Support HPCS...

Отзывы: