DMTA-10084-01EN, Rev. 2, November 2018
Calibration
91
Figure 8-1 UT dialog box
8.
If you are using an angle beam transducer, select the Angle icon (
), and then
press the Enter key to open the angle dialog box (see Figure 8-2 on page 92):
a)
Select
Angle
, and then use the adjustment knob to set the correct refracted
angle for the probe (
0
for a straight beam probe,
45
for a 45° probe, etc.).
b)
Press the Tab key to move to
Thick
, and then adjust the material thickness to
0.000
.
9.
Press the ESC key twice to return to the setup menu.
Содержание EPOCH 6LT
Страница 8: ...DMTA 10084 01EN Rev 2 November 2018 Table of Contents viii...
Страница 10: ...DMTA 10084 01EN Rev 2 November 2018 List of Abbreviations x...
Страница 16: ...DMTA 10084 01EN Rev 2 November 2018 Introduction 6...
Страница 20: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 1 10...
Страница 26: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 2 16...
Страница 32: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 3 22...
Страница 68: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 5 58 Figure 5 15 Diagnostic Test setup page...
Страница 109: ...DMTA 10084 01EN Rev 2 November 2018 Calibration 99 Figure 8 7 Calibration Successfully completed...
Страница 110: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 8 100...
Страница 154: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 9 144...
Страница 216: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 11 206 Figure 11 45 Corrosion sidebar menu G2Start...
Страница 220: ...DMTA 10084 01EN Rev 2 November 2018 Appendix 210...
Страница 226: ...DMTA 10084 01EN Rev 2 November 2018 List of Tables 216...
Страница 232: ...DMTA 10084 01EN Rev 2 November 2018 Index 222...