Chapter 11
232
If you need to continue capturing points, you can increase the instrument range or
increase the display delay to view echoes further out in time.
Once the correct number of points have been captured, choose
DAC/TCG > Done
to
complete the DAC curve and switch to the DAC inspection mode.
Figure 11-6 The completed DAC curve
Once the DAC curve is complete and the instrument is in inspection mode, the
instrument provides a new set of parameters:
•
DAC Gain
This adjustment allows you to manipulate the screen height/gain of both the
DAC/TCG curve as well as the echoes on screen. This allows for amplitude-to-
curve comparison at code compliant screen levels across the time base.
•
View
This parameter allows you to toggle between the DAC curve acquired and the
corresponding TCG setup based on the DAC curve information.
Содержание EPOCH 650
Страница 12: ...DMTA 10055 01EN Rev A February 2015 List of Abbreviations xii...
Страница 30: ...DMTA 10055 01EN Rev A February 2015 Introduction 18...
Страница 54: ...DMTA 10055 01EN Rev A February 2015 Chapter 1 42...
Страница 84: ...DMTA 10055 01EN Rev A February 2015 Chapter 3 72...
Страница 126: ...DMTA 10055 01EN Rev A February 2015 Chapter 5 114...
Страница 148: ...DMTA 10055 01EN Rev A February 2015 Chapter 7 136...
Страница 153: ...DMTA 10055 01EN Rev A February 2015 Programmable Inputs and Outputs 141 Figure 8 1 The A Out setup page...
Страница 154: ...DMTA 10055 01EN Rev A February 2015 Chapter 8 142...
Страница 236: ...DMTA 10055 01EN Rev A February 2015 Chapter 10 224...
Страница 332: ...DMTA 10055 01EN Rev A February 2015 Appendix B 320...
Страница 342: ...DMTA 10055 01EN Rev A February 2015 Appendix C 330...
Страница 346: ...DMTA 10055 01EN Rev A February 2015 Appendix D 334...
Страница 362: ...DMTA 10055 01EN Rev A February 2015 Index 350...