DMTA-10055-01EN, Rev. A, February 2015
Chapter 9
156
Figure 9-9 Entering the Velocity Cal thickness value
9.5
Calibrating on a Single Test Block of Known Thickness
You can also auto-calibrate on a single test block of known thickness. You can use
multiple back-wall echoes instead of coupling on both a thin step and a thick step. In
this scenario, leave the transducer coupled on the thin step, move the gate over to one
of the multiple back-wall echoes, and then enter the correct sound path thickness
(2, 3, 4, etc. multiple of the first back-wall echo) during the velocity portion of the
calibration.
9.6
Calibrating with a Dual Element Transducer
The sample dual element calibration described below is performed using an Olympus
transducer part number DHC711-RM with a frequency of 5.0 MHz and an element
diameter of 6 mm (0.25 in.).
Содержание EPOCH 650
Страница 12: ...DMTA 10055 01EN Rev A February 2015 List of Abbreviations xii...
Страница 30: ...DMTA 10055 01EN Rev A February 2015 Introduction 18...
Страница 54: ...DMTA 10055 01EN Rev A February 2015 Chapter 1 42...
Страница 84: ...DMTA 10055 01EN Rev A February 2015 Chapter 3 72...
Страница 126: ...DMTA 10055 01EN Rev A February 2015 Chapter 5 114...
Страница 148: ...DMTA 10055 01EN Rev A February 2015 Chapter 7 136...
Страница 153: ...DMTA 10055 01EN Rev A February 2015 Programmable Inputs and Outputs 141 Figure 8 1 The A Out setup page...
Страница 154: ...DMTA 10055 01EN Rev A February 2015 Chapter 8 142...
Страница 236: ...DMTA 10055 01EN Rev A February 2015 Chapter 10 224...
Страница 332: ...DMTA 10055 01EN Rev A February 2015 Appendix B 320...
Страница 342: ...DMTA 10055 01EN Rev A February 2015 Appendix C 330...
Страница 346: ...DMTA 10055 01EN Rev A February 2015 Appendix D 334...
Страница 362: ...DMTA 10055 01EN Rev A February 2015 Index 350...