![Olympus EPOCH 650 Скачать руководство пользователя страница 129](http://html2.mh-extra.com/html/olympus/epoch-650/epoch-650_user-manual_3929357129.webp)
DMTA-10055-01EN, Rev. A, February 2015
Managing Special Waveform Functions
117
Figure 6-2 Peak memory signal envelope example
This function is useful when it is necessary to find the peak from an indication during
an angle beam inspection.
The peak memory function cannot be activated in the unrectified
Rcvr > Rect =
RF
mode.
To activate peak memory function
1.
Press PEAK MEM.
The
symbol appears in the flag area to indicate that the function is active.
2.
Scan over a reflector to acquire the echo envelope.
3.
Press PEAK MEM again to turn off the peak memory function.
Peak memory flag
Peak memory signal envelope
(blue area)
Current live waveform (green
line)
Содержание EPOCH 650
Страница 12: ...DMTA 10055 01EN Rev A February 2015 List of Abbreviations xii...
Страница 30: ...DMTA 10055 01EN Rev A February 2015 Introduction 18...
Страница 54: ...DMTA 10055 01EN Rev A February 2015 Chapter 1 42...
Страница 84: ...DMTA 10055 01EN Rev A February 2015 Chapter 3 72...
Страница 126: ...DMTA 10055 01EN Rev A February 2015 Chapter 5 114...
Страница 148: ...DMTA 10055 01EN Rev A February 2015 Chapter 7 136...
Страница 153: ...DMTA 10055 01EN Rev A February 2015 Programmable Inputs and Outputs 141 Figure 8 1 The A Out setup page...
Страница 154: ...DMTA 10055 01EN Rev A February 2015 Chapter 8 142...
Страница 236: ...DMTA 10055 01EN Rev A February 2015 Chapter 10 224...
Страница 332: ...DMTA 10055 01EN Rev A February 2015 Appendix B 320...
Страница 342: ...DMTA 10055 01EN Rev A February 2015 Appendix C 330...
Страница 346: ...DMTA 10055 01EN Rev A February 2015 Appendix D 334...
Страница 362: ...DMTA 10055 01EN Rev A February 2015 Index 350...