105
5. Adjusting the Pulser/Receiver
This chapter describes how to adjust the EPOCH 650 ultrasonic flaw detector
pulser/receiver.
5.1
Adjusting the System Sensitivity (Gain)
To adjust the system sensitivity
1.
Press dB.
2.
Adjust the system sensitivity (gain) in either coarse or fine increments.
The total system sensitivity is 110 dB.
5.2
Using the AUTO XX% Feature
The AUTO XX% feature was formerly known as the AUTO-80 % in the EPOCH 4
Series. The default setting for AUTO XX% in the EPOCH 650 ultrasonic flaw detector
is 80 % full-screen height (FSH). You can adjust the FSH target value to meet the
demands of the application.
The AUTO XX% feature is used to quickly adjust the instrument gain (dB) setting to
bring the gated peak echo to XX% FSH. AUTO XX% is especially useful for bringing
the echo from a reference indication to XX% FSH to establish the instrument’s
reference gain level (see “Setting Reference Gain and Scanning Gain” on page 106 for
further information).
Содержание EPOCH 650
Страница 12: ...DMTA 10055 01EN Rev A February 2015 List of Abbreviations xii...
Страница 30: ...DMTA 10055 01EN Rev A February 2015 Introduction 18...
Страница 54: ...DMTA 10055 01EN Rev A February 2015 Chapter 1 42...
Страница 84: ...DMTA 10055 01EN Rev A February 2015 Chapter 3 72...
Страница 126: ...DMTA 10055 01EN Rev A February 2015 Chapter 5 114...
Страница 148: ...DMTA 10055 01EN Rev A February 2015 Chapter 7 136...
Страница 153: ...DMTA 10055 01EN Rev A February 2015 Programmable Inputs and Outputs 141 Figure 8 1 The A Out setup page...
Страница 154: ...DMTA 10055 01EN Rev A February 2015 Chapter 8 142...
Страница 236: ...DMTA 10055 01EN Rev A February 2015 Chapter 10 224...
Страница 332: ...DMTA 10055 01EN Rev A February 2015 Appendix B 320...
Страница 342: ...DMTA 10055 01EN Rev A February 2015 Appendix C 330...
Страница 346: ...DMTA 10055 01EN Rev A February 2015 Appendix D 334...
Страница 362: ...DMTA 10055 01EN Rev A February 2015 Index 350...