5.
Reinstall the outer probe-holder arm:
a)
Engage the outer probe-holder-arm pin in the other mounting hole in the
wedge.
b)
Ensure that the outer probe-holder arm is parallel to the inner probe-holder
arm.
c)
Tighten the outer probe-holder-arm screw.
6.
Ensure that the probe and wedge assembly is free to rotate around the pins
without presenting lateral movement.
2.4
Adjusting the Length of Exposed Cables and Tubes
After replacing a probe or reconfiguring the COBRA scanner for another pipe size,
you need to adjust the length of exposed cables and tubes.
To prevent damaging the probe and encoder cable, always carefully adjust their
length to fit the circumference of the scanner posts.
You do not need to remove the sleeve clamp to adjust the length of exposed cables and
tubes.
To adjust the length of exposed cables and tubes
1.
To gain more exposed cables and tubes length at the scanner end, hold the free
end of the protective sleeve, and pull the cables and the tubes through the
protective sleeve from the scanner end.
2.
To remove excess length at the scanner end, hold the tail of the scanner, and pull
the cables and the tube through the protective sleeve.
3.
On the scanner, clip the cables and the tube into the cable-management posts (see
Figure 2-6 on page 27).
Содержание COBRA
Страница 6: ...DMTA 20006 01EN Rev B February 2014 List of Abbreviations vi...
Страница 18: ...DMTA 20006 01EN Rev B February 2014 Important Information Please Read Before Use 12...
Страница 56: ...DMTA 20006 01EN Rev B February 2014 Chapter 3 50...
Страница 66: ...DMTA 20006 01EN Rev B February 2014 Chapter 4 60...
Страница 76: ...DMTA 20006 01EN Rev B February 2014 List of Tables 70...
Страница 80: ...DMTA 20006 01EN Rev B February 2014 Index 74...