Страница 1: ...REFLECTED LIGHT FLUORESCENCE ATTACHMENT CK40 RFA REPAIR MANUAL f OLYrvjPUS All rights reserved Reproduction in whole or in part without written permission is prohibited 8H f...

Страница 2: ...der unit C l 2 Cube dovetail diaphragm unit c 3 D REPAIR PROCEDURE 1 Adjustment procedure D l 2 Adjusting the cube working force D 2 3 Adjusting the click removing force of the cube D 2 4 Adjusting th...

Страница 3: ...is designed for B excitation and G excitation observation only This is 2 becausethe lamp housing U LHSOHG has a built in ultraviolet protective filter Use a shutter for transmitted light observation...

Страница 4: ...le dovetailtype with opticalpath click method 2 Maximum of 2 cubes 7 Shutter Slidetype with a click at thecenterandstopperson theboth ends 8 Filter With a filter pocket 32ND25 9 Reflectedlight 1 Suppo...

Страница 5: ...A OUTLINE OF PRODUCT CK40 RFA r 5 Dimensions Unit mm A 3...

Страница 6: ...4900 mN 2 980 InN using a tension gauge Optical axis 0 3 mm or less Combining the standard eyepiece KNO048 with adapter l a microscope stand product the standard objective KNOO15 and the cube BH2 UBF...

Страница 7: ...CK40 RFA C DISASSEMBLY AND ASSEMBLY PROCEDURES 1 Slider unit 0 Q C l...

Страница 8: ...CK40 RFA C DISASSEMBLY AND ASSEMBLY PROCEDURES pply adhesive to the screw c 2...

Страница 9: ...CK40 RFA C DISASSEMBLY AND ASSEMBLY PROCEDURES 2 Cube dovetail diaphragm unit q i a xiJ 1 c 3...

Страница 10: ...dhesive to the screws 3 Apply greaseto the dovetail surface only Refer to 5 Adjusting the optical axis 4 BASE FIXING SPRING OT2008 Apply greaseto the reverse side LEVER l OT1983 Apply adhesive to the...

Страница 11: ...ng the cube working force Refer to D 2 3 Adjusting the click removing force of the cube Refer to D 2 1 4 Adjusting the click removing force Of the lider I Refer to D 3 1 5 Adjusting the optical axis R...

Страница 12: ...esult is not within the standardrange adjustit by tighteningor looseningthefour pieces of SCREW 0 3 Adjusting the click removing force of the cube 3 1 Set the cube shown in the following illustration...

Страница 13: ...0 Insert a set of BALL SPRING 0 and SCREW move the SLIDER 0 to the clicking position of the SLIDER HOLDER l 0 and measure the click removing force using a TENSION GAUGE 0 Setting the end of the tensio...

Страница 14: ...om the LH MOUNT 0 of CKLCO RFA0 Instead of the LH MOUNT 0 mount the BXKC002 optical alignment jig and RFA unit on CK40 MAIN BODY 0 with AB3X25SA 2 Insert the CUBE CX DMB 2 0 into the dovetail of CK40...

Страница 15: ...the optical path 5 2 Setting the jigs shown below adjust the optical axis Centering telescop KN0029 Optical alignment jig for illumination part BXKC002 Standard objective for optical axis KN0015 5 3 T...

Страница 16: ...and move the CUBE DOVETAIL 0 to adjust the optical axis Standard Within 3 graduations on reticle scale of centering telescope KNOO29 If the result is not within the standard range return to Adjusting...

Страница 17: ...and products to check the FS image center l Eyepiece NCWHKlOX l Cube CX DMB 2 or CX DMG 2 product l Standard tube BKN0003 Use it when no observation tube is available l Reflected light fluorescence u...

Страница 18: ...the CK40 RFA on the CK40 main body turn off the power of main body set the microscope to reflected fluorescence observation and move the two KNOBS 0 to make sure that the FS image center can be set a...

Страница 19: ...X DMG 2 product OT1144 OT1145 KNO029 BXKC002 KNOOlS U LHSOHG product Tensiongauge 500 g Optical alignmentjig for illuminationpart Tensiongauge 1 kg Centeringtelescope D 2 D 2 3 D 4 6 1D 4 6 LB standar...

Страница 20: ...CHEMICALS f 1 List of lubricants Grease No Description OT1595 Siliconegrease OT2008 Grease medium c 2 4 c 4 Page 2 List of chemicals p Adhesive No OT1131 OT1983 Description Shellac Solvent based adhes...