Phenomena
Cause
Remedy
Page
Number
of beeps
CB
HS
MIX
3 times
–
–
The motorized
nosepiece is not
connected.
Set the main switch to (OFF) and
connect the motorized nosepiece
again.
9, 27
The motorized
nosepiece or its
cable is damaged.
Contact Olympus.
–
None
–
–
The hand switch
for motorized
nosepiece (BX3M-
HSRE) is not
connected.
Set the main switch to (OFF)
and connect the hand switch for
motorized nosepiece (BX3M-HSRE).
If the indicator4 on the side or the
back of the control box does not
turn ON even though the hand
switch for motorized nosepiece is
connected, contact Olympus.
9, 27
The hand switch
for motorized
nosepiece (BX3M-
HSRE) is damaged.
Contact Olympus.
–
None
–
–
The hand switch
for motorized
nosepiece
(BX3M-HSRE) or
the motorized
nosepiece is
damaged.
Contact Olympus.
–