Chapter 7 Cleaning, Disinfection, and Sterilization Procedures
113
ULTRASONIC BRONCHOFIBERVIDEOSCOPE BF-UC180F
Presoak for excessive bleeding and/or delayed reprocessing
Follow the steps below only in case of excessive bleeding
and/or delayed reprocessing: unnecessary immersions
should be avoided. Consecutive extended immersions may
damage the endoscope.
1.
Fill a basin with fresh detergent solution at a temperature and concentration
recommended by the detergent manufacturer. Use a basin that is at least
40 cm by 40 cm (16” by 16”) in size and deep enough to allow the
endoscope to be completely immersed.
2.
Carefully coil the endoscope’s insertion section and universal cord, and
completely immerse the endoscope in the detergent solution.
Do not coil the insertion section, universal cord and ultrasonic
cable with a diameter of less than 12 cm. Equipment damage
can result.
3.
Connect the suction cleaning adapter to the endoscope. Complete immerse
the endoscope in the detergent solution.
4.
Attach a 30 cm
3
(30 ml) syringe to the suction cleaning adapter. Withdraw
the plunger of the syringe to fill the instrument/suction channel and suction
cleaning adapter with the detergent solution (see Figure 7.21).
5.
Connect the syringe to the irrigation port and pull the plunger to fill the
balloon channel with detergent solution.
6.
With the endoscope, suction cleaning adapter and syringe completely
immersed in the detergent solution, disconnect the suction cleaning adapter
and syringe.
To prevent damage, do not pull the tubes of the suction
cleaning adapter when disconnecting it from the endoscope.
Always pull the caps.
7.
Soak the endoscope for 1 hour at the temperature and concentration
recommended by the detergent manufacturer.
8.
Remove the endoscope from the detergent solution.
9.
After soaking the endoscope, manually clean it following the procedures
described in this section.