DMTA-10022-01EN, Rev. C, January 2015
Software Options
81
Figure 6-9 Examples of the rectification modes
The available rectification modes are:
FULL
This mode shows the negative portion of the echo folded around the baseline, so
that both positive and negative waveform lobes are displayed. This mode
provides the best overall representation of position and magnitude for most
thickness-measurement applications.
FULL
is the default mode for dual element
transducers.
HALF
−
(
NEG
indicator)
This mode shows negative waveform lobes as positive, and does not show the
positive lobes.
HALF+
(
POS
indicator)
This mode shows positive waveform lobes, and does not show the negative
waveform lobes.
RF
This mode shows negative and positive lobes on either side of the baseline.
RF
is
the default mode for single element transducers.
Full
Half
−
Half+
RF
Содержание 45MG
Страница 8: ...DMTA 10022 01EN Rev C January 2015 Table of Contents viii List of Tables 227 Index 229...
Страница 26: ...DMTA 10022 01EN Rev C January 2015 Introduction 16...
Страница 36: ...DMTA 10022 01EN Rev C January 2015 Chapter 1 26...
Страница 42: ...DMTA 10022 01EN Rev C January 2015 Chapter 2 32...
Страница 136: ...DMTA 10022 01EN Rev C January 2015 Chapter 6 126...
Страница 150: ...DMTA 10022 01EN Rev C January 2015 Chapter 7 140...
Страница 157: ...DMTA 10022 01EN Rev C January 2015 Configuring the Instrument 147 Figure 8 3 The COMM screen...
Страница 158: ...DMTA 10022 01EN Rev C January 2015 Chapter 8 148...
Страница 196: ...DMTA 10022 01EN Rev C January 2015 Chapter 10 186...
Страница 208: ...DMTA 10022 01EN Rev C January 2015 Chapter 11 198...
Страница 226: ...DMTA 10022 01EN Rev C January 2015 Appendix A 216...
Страница 232: ...DMTA 10022 01EN Rev C January 2015 Appendix C 222...