DMTA-10022-01EN, Rev. C, January 2015
Chapter 4
50
4.6
Changing the Thickness Resolution
It is possible to change the thickness-measurement resolution, meaning the number of
digits shown to the right of the decimal point. The resolution selection affects all
displays and data output of values with thickness units. This includes measured
thickness, differential reference value, and alarm set points. The highest thickness
resolution with a dual element transducer is 0.01 mm (or 0.001 in.). The velocity is
always reported with full four-digit resolution.
The resolution can be reduced in certain applications where the extra precision of the
last digit is not required, or where extremely rough outside or inside surfaces make
the last thickness display digit unreliable.
The Single Element including High Resolution software option (P/N: 45MG-SE
[U8147022]) makes it possible to increase the resolution to 0.001 mm (or 0.0001 in.).
High resolution is available for measured thickness smaller than 102 mm (4 in.). High
resolution is not active with low-frequency transducers, or when the High Penetration
software option is activated.
To change the thickness-measurement resolution
1.
In the measurement screen, press
[SETUP]
, and then select
MEAS
.
2.
In the
MEAS
screen (see Figure 8-1 on page 142), set
RESOLUTION
to the
desired option:
•
STANDARD
: 0.01 mm (or 0.001 in.) [default]
•
LOW
: 0.1 mm (or 0.01 in.)
•
Optional
HIGH
: 0.001 mm (or 0.0001 in.)
3.
Press
[MEAS]
to return to the measurement screen.
Содержание 45MG
Страница 8: ...DMTA 10022 01EN Rev C January 2015 Table of Contents viii List of Tables 227 Index 229...
Страница 26: ...DMTA 10022 01EN Rev C January 2015 Introduction 16...
Страница 36: ...DMTA 10022 01EN Rev C January 2015 Chapter 1 26...
Страница 42: ...DMTA 10022 01EN Rev C January 2015 Chapter 2 32...
Страница 136: ...DMTA 10022 01EN Rev C January 2015 Chapter 6 126...
Страница 150: ...DMTA 10022 01EN Rev C January 2015 Chapter 7 140...
Страница 157: ...DMTA 10022 01EN Rev C January 2015 Configuring the Instrument 147 Figure 8 3 The COMM screen...
Страница 158: ...DMTA 10022 01EN Rev C January 2015 Chapter 8 148...
Страница 196: ...DMTA 10022 01EN Rev C January 2015 Chapter 10 186...
Страница 208: ...DMTA 10022 01EN Rev C January 2015 Chapter 11 198...
Страница 226: ...DMTA 10022 01EN Rev C January 2015 Appendix A 216...
Страница 232: ...DMTA 10022 01EN Rev C January 2015 Appendix C 222...