DMTA-10004-01EN, Rev. D, November 2016
Basic Operation
83
2.
In the menu, select the desired echo detection mode (
STANDARD
,
AUTO
E
‑
TO
‑
E
, or
MANUAL E
‑
TO
‑
E
).
3.
To perform the zero calibration again:
a)
Place a drop of couplant on the surface of the thin part of the test block.
b)
Couple the transducer to the thin part of the test block, and then press
[CAL ZERO]
.
c)
When the thickness reading is stable, press
[ENTER]
.
d)
Using the arrow keys, edit the thickness value to match the known thin
thickness of the test block.
5.6.1
Blanking Adjustments in Manual Echo-to-Echo Detection Mode
The 38DL PLUS offers two blanking functions to help detect valid echoes in situations
where material conditions generate unwanted signals:
EXT BLANK
The extended blank creates a blanked zone that begins at the left edge of the
waveform display and in which no signals are detected. In situations where the
second or third pair of back-wall echoes are stronger or cleaner than the first pair,
use the extended blank to control which pair of echoes to use for measurement.
E1 BLANK
The echo 1 (E1) blank runs for a selected interval following the first detected echo.
Use the E1 blank to exclude any unwanted peaks occurring between the first and
the second back-wall echoes. Unwanted peaks may be trailing edges of a large
first echo, or shear wave reflections on thick test pieces. The E1 blank parameter is
only available in the manual echo-to-echo detection mode.
To adjust the extended and the E1 blank parameters
1.
Select the manual echo-to-echo mode:
a)
Press
[2nd F]
,
[ZOOM] (E
‑
TO
‑
E)
.
b)
In the menu, select
MANUAL E
‑
TO
‑
E
.
2.
Press
[WAVE ADJ]
.
The wave adjustment parameter appears (see Figure 5-15 on page 84).
Содержание 38DL PLUS
Страница 9: ...DMTA 10004 01EN Rev D November 2016 Table of Contents ix Index 297 ...
Страница 10: ...DMTA 10004 01EN Rev D November 2016 Table of Contents x ...
Страница 28: ...DMTA 10004 01EN Rev D November 2016 Important Information Please Read Before Use 16 ...
Страница 30: ...DMTA 10004 01EN Rev D November 2016 Introduction 18 ...
Страница 50: ...DMTA 10004 01EN Rev D November 2016 Chapter 2 38 ...
Страница 60: ...DMTA 10004 01EN Rev D November 2016 Chapter 3 48 ...
Страница 102: ...DMTA 10004 01EN Rev D November 2016 Chapter 6 90 ...
Страница 221: ...DMTA 10004 01EN Rev D November 2016 Using the Datalogger 209 f Press MEAS to exit the minimum review ...
Страница 222: ...DMTA 10004 01EN Rev D November 2016 Chapter 11 210 ...
Страница 254: ...DMTA 10004 01EN Rev D November 2016 Chapter 13 242 ...
Страница 274: ...DMTA 10004 01EN Rev D November 2016 Chapter 14 262 ...
Страница 286: ...DMTA 10004 01EN Rev D November 2016 Chapter 15 274 ...
Страница 294: ...DMTA 10004 01EN Rev D November 2016 Appendix A 282 ...
Страница 300: ...DMTA 10004 01EN Rev D November 2016 Appendix C 288 ...
Страница 306: ...DMTA 10004 01EN Rev D November 2016 List of Figures 294 ...