DMTA-10004-01EN, Rev. D, November 2016
Dual Element Transducer Setups
215
b)
In the
SAVE TO
list, select the custom setup location where to save the setup.
c)
Select
SAVE
.
9.
Back in the
ACTIVE
screen, press
[MEAS]
to return to the measurement screen.
12.3 Recalling Stored Dual Element Transducer Setups
The 38DL PLUS allows you to store and recall setups for dual element transducers.
Stored dual element transducer setups can also include calibration information like
the velocity, the zero offset, the V-path, and the default gain. This feature makes it
easy to change between different transducer setups designed for different
applications.
Unsaved data will be lost when a different application setup is recalled. Save all
changes made to the active setup before recalling another setup.
To recall a custom dual element transducer setup
1.
Press
[XDCR RECALL]
.
2.
In the menu, select
CUSTOM DUAL ELEMENT
.
3.
In the
CUSTOM DUAL ELEMENT
screen, select the desired custom setup.
4.
In the
ACTIVE
screen, review the setup parameters.
5.
Press
[MEAS]
to return to the measurement screen with the recalled setup active.
12.4 V-Path
The V-path is the path along which the ultrasonic wave travels from one element to
the other when using a dual element transducer. Since the V-path is slightly longer
than twice the thickness, the instrument must calculate a compensation to measure
the true thickness. Using a V-path correction allows the 38DL PLUS to measure
thicknesses with the highest degree of accuracy and repeatability.
CAUTION
Содержание 38DL PLUS
Страница 9: ...DMTA 10004 01EN Rev D November 2016 Table of Contents ix Index 297 ...
Страница 10: ...DMTA 10004 01EN Rev D November 2016 Table of Contents x ...
Страница 28: ...DMTA 10004 01EN Rev D November 2016 Important Information Please Read Before Use 16 ...
Страница 30: ...DMTA 10004 01EN Rev D November 2016 Introduction 18 ...
Страница 50: ...DMTA 10004 01EN Rev D November 2016 Chapter 2 38 ...
Страница 60: ...DMTA 10004 01EN Rev D November 2016 Chapter 3 48 ...
Страница 102: ...DMTA 10004 01EN Rev D November 2016 Chapter 6 90 ...
Страница 221: ...DMTA 10004 01EN Rev D November 2016 Using the Datalogger 209 f Press MEAS to exit the minimum review ...
Страница 222: ...DMTA 10004 01EN Rev D November 2016 Chapter 11 210 ...
Страница 254: ...DMTA 10004 01EN Rev D November 2016 Chapter 13 242 ...
Страница 274: ...DMTA 10004 01EN Rev D November 2016 Chapter 14 262 ...
Страница 286: ...DMTA 10004 01EN Rev D November 2016 Chapter 15 274 ...
Страница 294: ...DMTA 10004 01EN Rev D November 2016 Appendix A 282 ...
Страница 300: ...DMTA 10004 01EN Rev D November 2016 Appendix C 288 ...
Страница 306: ...DMTA 10004 01EN Rev D November 2016 List of Figures 294 ...