- Installation Requirements
- Pre-Installation
The pre-installation is the suitability of the environment for the installation of the equipment. This step
should be guided by Olsen authorized assistance to ensure that the installation environment is suitable to receive
the new equipment. In this step should be prepared all water, compressed air, sewage tank, power supply and
other accessories that will interact with the equipment.
- Water for the Water Tank
The ideal pressure for the equipment water supply is 2.8 to 6.0 bar.
The recommended pH (Potential of Hydrogen) is between 6.5 and 8.
The water used in the tank must be filtered and potable. It is recommended to use mineral water, which
can easily be obtained in any market, but also can use public water, provided it is filtered and boiled, to avoid
possible debris in the water causing valves clogging.
- Electrical Installation
The power grid must be single-phase, has specific grounding and exclusive 10 A/30 mA DR circuit breaker.
The circuit breaker must only supply the dental equipment and must be easy and quick to disconnect from the
mains.
If the power grid presents voltage variation, the installation of a surge protection device is required. There
is a table below for sizing the electrical installation:
Before turning on the equipment make sure that the electrical voltage of the equipment is the same
voltage as the power grid. There is a risk of damage to the equipment!
The contact with clean and filtered water offers no known risk to the user or the patient, but the use of
sanitized gloves is recommended to fill the tanks and avoid cross-contamination.
This equipment should only be connected to a power source with protective grounding. There is an
electric shock risk.
Do not place the equipment on a place that makes it difficult to disconnect it from the power grid.
Do not connect this equipment to a power grid without a 30 mA DR circuit breaker.
An authorized technician should make the installation. Installation performed by an unauthorized
person will result in warranty loss.
Do not connect any other equipment on the cart equipment’s exclusive power supply
.
8
- Installation
The installation consists of equipment assembling and installation of its supply connections, such as water,
compressed air, sewage (drainage and vacuum when available) and electricity.
When assembling the equipment, the technician should make the pressure adjustment of the terminals
for pneumatic turbines, adjustment of the water pressure and air of the syringe and spray of pneumatic
instruments, among other activities.
The equipment installation must be performed by Olsen certified technician, who will also inspect the
equipment making sure it's according to what was requested, if it remains intact after having been transported,
in addition, to performing a quality inspection before use, ending the process with a guidance on proper
equipment operation, cleaning, and maintenance.
Voltage (V) Cable Size (mm²) Distance (m)
Electric Current (A)
127/220
2,5
Up to 20
10