WX16 User Manual | 64 

7.5.4 Coax 

Cables 

If a coax cable connects to the antenna via connectors, it is very important to 
weatherproof the connection using our 1000-2314 or equivalent sealing tape. Moisture 
ingress into a coax cable connection is the most common cause of problems with 
antenna installations. A three layer sealing process is recommended – an initial layer 
of electrical PVC tape, followed by a second layer of self-vulcanizing weatherproofing 
tape (1000-2314), with a final layer of electrical PVC tape. 

Allowing a drip “U loop” of cable before the connection is also a good idea. The loop 
allows water to drip off the bottom of the U instead of into the connection, reduces 
installation strain and provides spare cable length in case later the original connectors 
need to be removed, the cable cut back and new connectors fitted. 

Avoid installing coax cables together in long parallel paths. Leakage from one cable to 
another has a similar effect as mounting an antenna near another antenna. 

7.6 

Surge Protection & Grounding 

Voltage surges can enter the WX16 via the antenna connection, power supply 
connection, connections to other equipment and even the “earth” or “ground” 
connection. Surges are electrical energy following a path to earth, and the best 
protection is achieved by “draining” the surge energy to earth via an alternate path. 
Wireless devices need to have a solid connection to earth via a ground stake or ground 
grid if the soil has poor conductivity. Solid connection means a large capacity 
conductor (not a small wire) with no coils or sharp bends. All other devices connected 
to the WX16 need to be grounded to the same ground point. There can be significant 
resistance between different ground points leading to very large voltage differences 
during lightning activity. As many wireless units are damaged by earth potential surges 
due to incorrect grounding as are damaged bydirect surge voltage. 

It is very difficult to protect against direct lightning strikes, but the probability of a direct 
strike at any one location is very small. Unfortunately, power line surges and 
electromagnetic energy in the air can induce high voltage surges from lightning activity 
several miles away. 

7.6.1 Antenna 

Grounding 

Electromagnetic energy in the air will be drained to ground via any and every earth 
path. An earth path exists between the antenna and the WX16, and to protect against 
damage, this earth path current must be kept as small as possible. This is achieved by 
providing better alternate earth paths. It is important to ground the antenna to the same 
ground point as the WX16. Antennas are normally mounted to a metal bracket that 
should be grounded to the WX16 earth connection. Surge energy induced into the 
antenna will be drained first by the mount’s ground connection, second by the outside 
shield of the coax cable to the ground connection on the radio and third by the internal 
conductor of the coax cable via the radio electronics. This third earth path causes 

Содержание WX16

Страница 1: ...WX16 Sixteen Channel Controller User Manual P N 77036022 1 Revision 0 Reference Firmware 6 09 rev 4 7...

Страница 2: ......

Страница 3: ...r Industrial Scientific Oldham nor any other associated company can be held liable for any damages including without limitations damages for the loss or interruption of manufacture loss of information...

Страница 4: ...the Keypad 10 2 2 Channel Configuration Menus 11 2 2 1 Channel Setup Entry Menu 12 2 2 2 Alarm 1 Alarm 2 Horn Relay Setup Menu 12 2 2 3 Alarm 3 Fault Alarm Menu 13 2 2 4 Data From Menu to Set Input So...

Страница 5: ...upply 40 SECTION 4 41 4 0 System Diagnostics 41 SECTION 5 44 5 0 ModBus RS 485 Ports 44 5 1 ModBus Slave Register Locations 44 SECTION 6 52 6 1 Panel Rack Mount Enclosure 52 6 2 NEMA 4X Wall Mount Fib...

Страница 6: ...rtions of this instrument have been assessed by CSA for C22 2 No 152 performance requirements This equipment is suitable for use in Class I Division 2 Groups A B C and D or non hazardous locations onl...

Страница 7: ...master RS 485 port A ModBus RTU slave RS 485 port is also standard for sending data to PC s PLC s DCS s or even other WX16 Controllers Options such as analog I O and discrete relays for each alarm are...

Страница 8: ...trend screen indicates channel real time reading in engineering units measurement name range and MIN MAX AVERAGE values for the preceding 24 hour period The SI field on the top right indicates number...

Страница 9: ...devices such as relays lamps or transmitters Power consumed from these terminals should be considered when calculating system power consumption 1 2 1a 150 Watt AC 24 VDC Power Supply 110 120 VAC 3 2A...

Страница 10: ...ic keypad 1 2 6a Non Intrusive Magnetic Keypad The WX16 operator interface includes five front panel touch keys A magnetic keypad option offers these five keys with adjacent magnetic keys This option...

Страница 11: ...be displayed in sequence with the UP DOWN keys The NEXT key switches between the 3 graphic data screens When WX16 power is applied the graphic LCD returns to the screen active when power was last rem...

Страница 12: ...inter controlled by the UP DOWN keys indicates the selected variable Some are simple YES NO or ON OFF entries toggled by pressing the EDIT key Others such as Measurement Name and Eunits fields may hav...

Страница 13: ...nel Configuration Menus Figure 2 1 illustrates the menu tree for configuring Channel variables These items affect only the specific channel selected System specific variables are in the menu tree show...

Страница 14: ...s for this alarm to automatically reset as soon as the alarm condition clears TRIP ON is set to HIGH for increasing alarms or LOW for decreasing alarms to determine if the alarm activates upon exceedi...

Страница 15: ...ip point entered prior to changing the menu to Alarm 3 The following example describes how to configure both the Fault out of range and Alarm 3 level trip points for a channel Example If the common Fa...

Страница 16: ...t or Analog with Local Cal is selected These selections should only be used when the input originates from a non calibrated signal source such as the Bridge Sensor Input option described in section 3...

Страница 17: ...vices providing WX16 inputs also indicate special modes of operation such as Calibration Maintenance or Fault by transmitting a special 4mA or negative Marker value The WX16 offers channel Marker menu...

Страница 18: ...and records it as the channel s Sensor Life Sensor Life is stored in the WX16 ModBus database and displayed as a bar graph in the Sensor Info screen see section 2 3 6 It is a useful tool for planning...

Страница 19: ...nit and 16 character Measurement Name ASCII fields Eunits should define the units of measure for what this channel is to display Measurement Name should describe the source of this data in the user s...

Страница 20: ...tilized Turning it off will cause the controller to never process inputs applied to this channel and no alarms will be tripped or data displayed Inactive channels have a line drawn through them on the...

Страница 21: ...again at later dates when degradation may have occurred To check zero calibration apply the ZERO calibration value to the sensor and observe the live reading If the zero reading differs from the zero...

Страница 22: ...ems needing configuration are not specific to a channel but affect the entire WX16 system These are located in the system entry menu shown on the left side of Figure 2 10 System menus are accessed by...

Страница 23: ...elay 1 while channels 9 16 trip relays on an optional discrete relay PCB configured for Alarm 1 see section 3 2 Failsafe controls relay activation for this common relay Failsafe ON causes the relay to...

Страница 24: ...on in a Failsafe mode using the System Setup menu shown in Figure 2 12 Entering YES causes these discrete relays to have energized coils when no alarm condition exists for the associated channel and d...

Страница 25: ...ng Acknowledge OFF allows the common Horn relay to drive devices other than horns or sirens such as a light or a fan Display Alm YES forces the LCD to display the Bar Graphs screen upon any new alarm...

Страница 26: ...oils of slave devices are unaffected LCL only writes to the 2001 coils of connected RTU s when the local Reset pushbutton is pressed MB LCL writes these coils if this WX16 s 2001 coil has been reset v...

Страница 27: ...nfo screen displays each channel s sensor status as illustrated in Figure 2 16 Channels with Sensor Life disabled indicate Option Disabled above the corresponding empty bar graph If Sensor Life is ena...

Страница 28: ...following procedure Remove power from the controller Reapply power and as the alarm LED s begin scrolling down hold the following keys simultaneously UP DOWN NEXT EDIT Watch closely The 4 digit autho...

Страница 29: ...s the I O PCB to the WX16 CPU and Display nest assembly Two I 2 C bus connectors allow addition of optional functions such as analog I O and discrete alarm relays for each channel Horizontal jumpers i...

Страница 30: ...after power up indicating what options are connected and for which channels This information is also available from the Diagnostics Mode described in Section 4 Figure 3 2 3 1 1 Optional Analog Input P...

Страница 31: ...re to be applied EXC and IN LO terminals are bussed together internally EXC terminals are tied directly to TB2 1 and IN LO terminals are tied to TB2 2 Bussing allows transmitter power to be brought in...

Страница 32: ...e configured via rotary switch S1 to function for ALARM 1 ALARM 2 or ALARM 3 FAULT for channels 1 8 or 9 16 A 1 minute time delay after power is provided to inhibit relay actuation until the system ha...

Страница 33: ...e limited in ability to carry this power further than a few feet without a significant voltage drop Some WX16 applications with relays for all 48 alarms may require up to 6 boards TB2 allows a heavier...

Страница 34: ...hannel 10 0191 Each 10 0192 channel is equipped with a bridge amplifier and balance potentiometer and an adjustable switching regulator for setting the correct sensor excitation voltage A 3 position c...

Страница 35: ...set the Voltage Adjust potentiometers for the correct sensor excitation voltage This may range from 1 5 volts to 7 5 volts depending upon sensor specifications Sensors may be damaged by accidental ov...

Страница 36: ...TD Sensor Input board shown in Figure 3 6 allows these popular temperature sensors to be connected directly to the WX16 without additional signal conditioning or transmitters A 3 position range jumper...

Страница 37: ...sensor readings Each channel must be calibrated individually by either simulating desired zero and span calibration resistance values or by actually placing the channel s RTD into an actual precision...

Страница 38: ...WX16 User Manual 36 Figure 3 6...

Страница 39: ...supply With at least 20 volts DC primary power they are capable of driving 20mA through a 750 ohm load Outputs are self powered and DC power should not be provided by the receiving device Note This P...

Страница 40: ...data bits no parity one stop bit and FlowControl Hardware Printer diagnostic red LED s indicate printer faults such as out of paper overflowed buffer or loss of communications Green LEDs flicker to c...

Страница 41: ...w setting of correct local time and date The ALARM PRINT ON OFF entry allows printing to be discontinued if turned to OFF PORT allows selection of RS 232 RS 422 or the parallel port With only one port...

Страница 42: ...r for the monitors or transmitters providing inputs to the WX16 A 150 watt AC DC power supply may be included for these applications 115 VAC or 230 VAC selected via slide switch When ordered from the...

Страница 43: ...nd analog I O It is exited manually by pressing NEXT and automatically if no keys are pressed for 5 minutes It is very important to understand that CHANNEL INPUT DATA IS NOT PROCESSED DURING THE DIAGN...

Страница 44: ...WX16 User Manual 42 Figure 4 1...

Страница 45: ...WX16 User Manual 43 Figure 4 2...

Страница 46: ...Notes Channel Tag 1 40401 40408 3 n a 2 characters per register Channel Tag 2 40409 40416 3 n a 2 characters per register Channel Tag 3 40417 40424 3 n a 2 characters per register Channel Tag 4 40425...

Страница 47: ...547 40549 3 n a 2 characters per register 3 registers per channel EUNITS 8 40550 40552 3 n a 2 characters per register 3 registers per channel EUNITS 9 40553 40555 3 n a 2 characters per register 3 re...

Страница 48: ...95 40597 3 n a 2 characters per register 3 registers per channel ASCII Value 8 40598 40600 3 n a 2 characters per register 3 registers per channel ASCII Value 9 40601 40603 3 n a 2 characters per regi...

Страница 49: ...nd status bits for a channel They are as follows Type Channel First Last Read FC Write FC Notes Channel Status 1 16 31017 31032 4 n a 16bit integer see bit by bit definition below Alarm 1 Trip bit0 1...

Страница 50: ...Write FC Notes Cal Status 1 16 31037 n a 4 n a packed 16bit integer Trend Interval 1 16 31038 n a 4 n a 16bit integer Timer Time in Seconds Fault Status 1 16 31039 n a 4 n a packed 16bit integer Alar...

Страница 51: ...integer Sensor Life 9 31073 n a 4 n a Signed 16bit integer Sensor Life 10 31074 n a 4 n a Signed 16bit integer Sensor Life 11 31075 n a 4 n a Signed 16bit integer Sensor Life 12 31076 n a 4 n a Signe...

Страница 52: ...of 1 2 3 or 4 where 123 4 would return the value 10 Type Channel First Last Read FC Write FC Notes Zero Real 1 16 41001 16 n a 3 n a zero real w o decimal point Zero DP 1 16 41017 32 n a 3 n a zero re...

Страница 53: ...om the ModBus master Type Channel First Last Read FC Write FC Notes Update MIN n a 2065 n a n a 5 Moves 24 hour MIN data trend to trend data base Update AVG n a 2066 n a n a 5 Moves 24 hour MIN data t...

Страница 54: ...in a rectangular cutout Only two 8 channel I O option PCB s such as analog input or discrete relays may be mounted directly to the back of the enclosure Additional 8 channel I O option PCB s must be l...

Страница 55: ...ve mounted weather deflector shield is recommended Weight is approximately 17 pounds Figure 6 3 provides important warning information concerning correct grounding procedures for non metallic enclosur...

Страница 56: ...WX16 User Manual 54 Figure 6 3...

Страница 57: ...nside this enclosure It is suitable for mounting outdoors but an above mounted weather deflector shield is recommended Weight is approximately 30 pounds Conduit entries are not provided so installers...

Страница 58: ...osure and follow national and local electrical codes Figure 6 5 6 5 Main I O Option PCB Footprint Dimensions WX16 controllers have virtually unlimited possibilities for configuration of options such a...

Страница 59: ...expansion plate If more than 7 option positions are required the 10 0178 NEMA 4X wall mount option enclosure supporting 8 positions may be added WX16XP NEMA 7 Wall Mount supports 5 option positions a...

Страница 60: ...connected to the WX16 s RS 485 master port and Wireless ModBus Slave mode connects it to the RS 485 slave port Figure 3 1 It is important to remember RADIO SETUP functions described in section 7 1 may...

Страница 61: ...go InRange of or communicate with a transceiver operating on a different Hop Channel System ID may be set from 1 255 using the WX16 keypad and is similar to a password character or network number and...

Страница 62: ...the WIRELESS RF900 setting shown at center of Figure 7 3 ONLY FOR COMMUNICATION TO OLCT 200 WIRELESS TRANSCEIVERS See section 7 4 for setting up wireless networks with other ModBus slave devices OLCT...

Страница 63: ...ndicates the OLCT 200 s integral 3 6V lithium D cell part 10 0293 has dropped to below 3 3V and should be replaced very soon LoBatt is indicated on the WX16 s LCD readout and the Alarm 3 LED flashes R...

Страница 64: ...ribed in section 3 1 This is a popular application when the ModBus slave is another remote WX16 or Oldham s WX4 controller available with built in radio modem compatible with the WX16 Radio Kit Other...

Страница 65: ...st also have vertical polarity Care needs to be taken to ensure the Yagi is aligned correctly to achieve optimum performance Two Yagis can be used for a point to point link In this case they can be mo...

Страница 66: ...earth via a ground stake or ground grid if the soil has poor conductivity Solid connection means a large capacity conductor not a small wire with no coils or sharp bends All other devices connected t...

Страница 67: ...hout a surge diverter 7 6 2 Connections to Other Equipment Surges can enter the wireless unit from connected devices via I O serial or Ethernet connections Other data devices connected to the wireless...

Страница 68: ...LE EAST AFRICA Industrial Scientific Oldham Z I Est rue Orfila BP 20417 62 027 Arras Cedex France Tel 33 3 21 60 80 80 Fax 33 3 21 60 80 00 Email info oldhamgas com Website www oldhamgas com ASIA PACI...