24
2
1
3
Objective Band Heater
Velcro Tape
Figure 13. Objective Heater (OBJ-COLLAR ####) Assembly
8.
Connect the power cord to the Control Unit.
Once the system has been initialized, the UNO-COMBINED-CONTROLLER display will show its
homepage.
Tip
►
Wait until the system has reached the steady state before loading your samples.
Tip
►
To replace a broken fuse disconnect the unit from the power source.
Use a flat head
screw driver to replace the fuse if it is broken.
Incubation chamber
UNO-COMBINED-CONTROLLER must be equipped with an incubation chamber compatible with
you stage band/model. Select the chamber on
system is shipped from the factory already set according to the incubating chamber model and
sample adapter you have ordered. You can see the current base, lid and HM offsets on the
display (See paragraph 13.5.1).
Note
►
Thermal control settings are a function of the adapter in use. Hence, every time you
change adapter you must input the correct thermal settings using the procedure described in
paragraph 13.5.1. In case you purchase additional incubating chamber models make sure to
input the correct base offset in the thermal settings prior to starting your experiment. A
careful calibration has been performed in our laboratories for each incubation chamber and we
guarantee that sample temperature is maintained at the desired value by selecting the adapter
and the chamber in use. See Table 1, for values of base offset temperature. Input the correct
base offset in the Control Unit Setting using the push button display (See paragraph 13.5.1).