Heating /
Cooling Box
Condensate
Vessel
Condensate
Drainer
Figure 15. Condensate Drainer
–
assembly.
3.
Connect the Condensate Drainer to the main power supply using its supplied power cord.
7.5.1
How to configure the Condensate Drainer
The Condensate Drainer has the dosing quantity parameter set by default at 10 8H. This means that the
Condensate Drainer doses 10ml of water every 8 hours. Okolab suggest changing the parameters to 20 1H to dose
20 ml of water every hour. For setting the dosing quantity, press button no.1 (1 ml -steps) and/or button no.2 (10
ml
–
steps) as often as the desired dosing quantity is indicated. For setting the interval time, press button no.3 as
often as the desired value is indicated.
Note
►
If the system runs continuously in cooling mode, check the water content inside the Condensate Vessel
regularly. The Condensate Vessel should be emptied at least once a week. To empty the Condensate Vessel,
remove the green marked tube from the push to fit connector and unscrew the cup of the vessel. If the setpoint
temperature is higher than the Room temperature, there is no need to check the water level.
How to start up the system
To start up H201-T-UNIT-BL-CRYO, follow the instructions below:
Before starting up the Chiller, ensure that the bypass adjustment wheel is completely closed. (see
Chiller’s User Manual, pag. 16
or Figure 16). Turn the bypass adjustment wheel anticlockwise as far
as the stop.
1.
Switch the Chiller on.